AGL7 module adjustment procedures

AGL7 module adjustment procedures

Requirements for adjustment

Personnel
Warm-up period
Access

Required equipment

Table 4-1: Equipment required for AGL7 module adjustment procedures

Item
TV signal generator platform
Digital multimeter
Oscilloscope
75 Ω BNC cable
75 Ω coaxial terminator
BNC T connector
BNC female to dual banana
adapter
4–2
This section contains information needed to adjust the AGL7 module.
Adjustments in the AGL7 module are for the output offset, output gain, and
clock timing.
Before doing the adjustment, note the following requirements.
This procedure is only to be performed by trained service technicians.
The AGL7 module requires a 20 minute warm-up time in a +20 °C to +30 °C
environment before it is adjusted. Adjustments done before the operating
temperature has stabilized may cause errors in performance.
When adjusting the output offset, output gain, and clock timing, you need to
remove the top cover of the TG8000 mainframe. (See page 2-25, Top cover.)
The following table lists the equipment required to adjust the output offset, output
gain, and clock timing.
Minimum requirement
No.
1
1
5 1/2 digits
1
Bandwidth: 500 MHz or higher
1
Length: 42 inches
1
1
1
Recommended equipment
Tektronix TG8000
FLUKE 8842A
Tektronix DPO70404C with two P6139A
probes
Tektronix part number 012-0074-00
Tektronix part number 011-0102-03
Tektronix part number 103-0030-00
Tektronix part number 103-0090-00
TG8000 Multiformat Test Signal Generator Service Manual