Invacare® HomeFill® II Oxygen System

7 Troubleshooting

7.1 Indicator Lights

Indicator Light Color
None
O
BELOW NORMAL - YELLOW
2
FILLING - GREEN
FULL - GREEN
ATTENTION - RED (Audible Alarm)
28
Compressor Status
Compressor is Off.
Compressor is On and warming up (0-3
minutes after power is On).
O
from concentrator is < 90% (any time after
2
three minute warmup).
Compressor is On and cylinder is filling. O
from concentrator is >90% (any time after
three minute warmup).
Compressor is On. Cylinder is finished filling
(full) -remove cylinder (FULL light is On in
>10 mins.)
or
Cylinder is not attached properly - check
connection (FULL light is On in <10 mins.).
Compressor is On. Cylinder NOT filling.
Causes:
Full cylinder attached when unit turned On -
replace with empty cylinder
or
Internal failure - contact dealer.
Effect
O
flowing to user if concentrator is on.
2
Cylinder filling not started.
O
flowing to user if concentrator is On.
2
Cylinder filling stops, will restart when O
90%.
O
flowing to user if concentrator is On.
2
Cylinder is filling.
2
O
flowing to user if concentrator is On.
2
Cylinder filling stops.
Cannot fill another cylinder until compressor
is turned Off and On again.
O
flowing to user if concentrator is On.
2
Compressor does not begin to fill or stops
filling.
Unit must be turned Off and On before filling
can occur.
O
flowing to user if concentrator is On.
2
>
2
1145804-F-00