2.3 Dealer Information

Invacare recommends leaving a full cylinder of oxygen with
the patient after setting up the HomeFill compressor.
Before leaving a HomeFill Compressor with a user, the following
checklist MUST be completed:
q
Make sure the pressure test on cylinder has not expired.
q
Check the EMC section for operational information and possible
RF issues. Refer to EMC Information.
q
Instruct the user on the safe operation of the concentrator and
review all warnings.
q
Instruct the user on the safe operation of the HomeFill
compressor and review all warnings.
q
Leave a full cylinder of oxygen with the user.
q
Leave a copy of all warnings and filling instructions with the user.

2.4 Double-Insulated Products

CAUTION!
– DO NOT unplug the power cord by pulling on
the cord. To unplug, grasp the plug not the cord,
otherwise damage may occur.
In a double‐insulated product, two systems of insulation are
provided instead of grounding. No grounding means is required
on a double‐insulated product, nor is a means for grounding to be
added to the product. A double‐insulated product is marked with
the words "DOUBLE INSULATION" or "DOUBLE INSULATED"
or the symbol:
1145804-F-00

2.5 EMC Information

WARNING!
Risk of Injury or Damage
To reduce the risk of injury or product damage from
interference with wireless equipment:
– Keep the compressor at least 9.8 ft (3.0 m) away from
wireless communication equipment such as wireless
home network devices, mobile phones, cordless
phones and base stations, walkie-talkies, etc.
Medical Electrical Equipment needs to be installed and used according
to the EMC information in this manual.
This equipment has been tested and found to comply with EMC
limits specified by IEC/EN 60601-1-2 for Type BF equipment.
These limits are determined to provide reasonable protection
against electromagnetic interference in a typical home healthcare
environment. This equipment is not intended for use in a professional
healthcare environment.
Portable and mobile RF communications equipment can affect the
operation of this equipment. The compressor may be stacked on top
on an Invacare HomeFill compatible oxygen concentrator during
setup. The compressor operation should be observed to verify
normal cylinder filling in either situation.
Other devices may experience interference from even the low levels
of electromagnetic emissions permitted by the above standard. To
determine if the emissions from the compressor are causing the
interference, turn the compressor Off. If the interference with the
other device operation stops, then the compressor is causing the
interference. In such rare cases, interference may be reduced or
corrected by one of the following measures:
Safety
11