The Hydrostatic Testing Date

Aluminum cylinders MUST undergo testing every ten years.
WARNING!
Risk of Injury
To reduce the risk of injury:
– DO NOT fill cylinders that have not been tested in the past ten years. Otherwise, serious personal injury may result. Contact
your dealer/Home Care equipment provider for replacement.
External Examination
1.
Examine the outside of the cylinder for the following conditions, and replace the cylinder if they exist:
Dents or dings
Arc burns
Oil or grease
Any other signs of damage that might cause a cylinder to be unacceptable or unsafe for use.
2.
Examine the cylinder for evidence of fire or thermal damage. Evidence includes charring or blistering of the paint, or other protective coating
or heat sensitive indicator. If fire or thermal damage is found, replace the cylinder.
3.
Inspect the cylinder/regulator assembly for the following and replace if found:
Debris, oil or grease
Noticeable signs of damage
Signs of corrosion inside the valve
Signs of excessive heat or fire damage
1145804-F-00