7 Troubleshooting

7.1 Indicator Lights

Indicator Light Color
None
O
BELOW NORMAL - YELLOW
2
FILLING - GREEN
1100873-M
Compressor Status
Compressor is Off.
Compressor is On and warming up (0-3
minutes after power is On).
O
from concentrator is < 90% (any time
2
after three minute warmup).
Compressor is On and cylinder is filling.
O
from concentrator is >90% (any time
2
after three minute warmup).
Effect
O
flowing to user if concentrator is on.
2
Cylinder filling not started.
O
flowing to user if concentrator is On.
2
Cylinder filling stops, will restart when
O
> 90%.
2
O
flowing to user if concentrator is On.
2
Cylinder is filling.
O
flowing to user if concentrator is On.
2