Invacare® HomeFill® II Oxygen System
O
below normal (YELLOW) light stays On
2
Fill times exceed maximum as
recommended fill times OR GREEN filling light stays On.
Excessive noise when compressor is On
30
Problem
1.
Ensure the concentrator has warmed up for at least 20 minutes.
If it has not been 20 minutes, turn the compressor Off until the
concentrator warms up. If the YELLOW light still remains on
proceed to step 2.
2.
Inspect the connection between the compressor and the
concentrator for damage or pinching.
If damaged, replace.
If pinched, straighten tubing. Turn the compressor Off (O)
for at least 10 seconds. Turn the compressor On ( I ). If the
YELLOW light still remains on after 10 minutes, proceed to
step 3.
3.
The concentrator may need service, contact your home health
care provider or Invacare.
1.
Ensure the cylinder regulator is set to Off (O).
2.
Check the recommended fill times shown in the individual
cylinder's manual.
3.
Check that the cylinder is connected properly. Turn the
compressor Off (O). Remove the cylinder and reinstall. Turn the
compressor On ( I ).
4.
Check the compressor/concentrator line for proper connection
and/or damage (i.e. - leaks or kinks).
5.
Check the concentrator liter flow setting is at 2.5 L/min or less
for Platinum 5 or Perfecto
Platinum 9 concentrators. If necessary, adjust the liter flow setting
(refer to the concentrator User manual). If the GREEN filling light
does not go Off, the compressor and/or concentrator may need
service. Contact your home health care provider or Invacare.
1.
Call home health care provider or Invacare for service.
Solution
concentrators or 5 L/min or less for
2
1145804-F-00