SECTION 6—TROUBLESHOOTING/ CYLINDER FILL TIMES
SECTION 6—TROUBLESHOOTING/
CYLINDER FILL TIMES
PROBLEM
No indicator lights on
when compressor turned
on.
RED Light ON
Compressor cycles ON
and Off
or
O
BELOW NORMAL
2
(YELLOW) and FILLING
(GREEN)
lights are alternating
or
O
BELOW NORMAL
2
(YELLOW)
light goes On and Off
Part No 1100873
TROUBLESHOOTING
1. Turn compressor Off.
2. Check that power cord is properly plugged
into compressor and wall outlet
3. Turn the compressor On.
4. Clean air filters. Refer to Maintenance on
page 37.
5. If performing STEPS 1-4 does not cause any
indicator lights to come on, contact your home
health care provider or Invacare for service.
1.Check the cylinder gauge to see if it is FULL.
Cylinder full - Remove the cylinder and turn the
unit Off. Perform the Cylinder Prefill Inspection
on page 16 with a cylinder that is NOT FULL.
Cylinder not full - Proceed to STEP 2.
2. Check that the cylinder is connected properly.
Turn the compressor Off. Remove the cylinder
and reinstall. Turn the compressor On.
3. If cylinder is not full and is connected properly,
the compressor may have internal failure.
Contact your home health care provider or
Invacare.
1. Check the concentrator liter flow setting is at 3
L/min. or less.
If necessary, adjust the liter flow setting (refer to
the concentrator owner's manual). Wait at least
20 minutes for the GREEN light to appear. If it
does not appear, proceed to STEP 2.
2. The concentrator may need service, contact
your home health care provider or Invacare.
34
SOLUTION
.
HomeFill
II Compressor