Hach BioTector B3500C User Manual page 117

Online toc analyzer
Hide thumbs Also See for BioTector B3500C:
Table of Contents

Advertisement

SYSTEM PROGRAM 2 TC
SAMPLE IN
SYSTEM PROGRAM 2 VOC
SYSTEM PROGRAM 3
SYSTEM PROGRAM 3 TIC + TOC
RANGE CHANGE 2-3
RANGE CHANGE 3-2
RANGE CHANGE 3-1
SAMPLE IN
TIC ACID
BASE
BASE OXIDATION
TOC ACID
5p , P-V
 SAMPLE IN defines the amount of sample (in pulses) to be
 All system parameters for SYSTEM PROGRAM 2 VOC analysis
150.0 mgC/l
 RANGE CHANGE 2-3 defines the concentration level for the
 The RANGE CHANGE 2-3 is typically the top concentration
 RANGE CHANGE 3-2 specifies the concentration level for the
120.0mgC/l
 The RANGE CHANGE 3-2 is typically 20% less of the top
 RANGE CHANGE 3-1 defines the concentration level for the
40.0mgC/l
 The RANGE CHANGE 3-1 is typically 20% less of the top
 SAMPLE IN states the amount of sample (in pulses) to be
3p
P-V
12p
 TIC ACID defines the amount of acid reagent (in pulses) to be
 BASE specifies the amount of base (in pulses) to be injected into
70p
130s
 BASE OXIDATION defines the oxidation time (150s by default),
 If the Base Oxidation time is different than the one programmed
 TOC ACID states the amount of acid (in pulses) to be injected
58p
injected into the reactor in TC mode at range 2.
mode are covered in System Program 2 TC and System
Program 2 TIC + TOC above.
BioTector to change its range automatically from range 2 to
range 3.
point of TOC RANGE 2 in System Range Data menu by default.
BioTector to change its range automatically from range 3 to
range 2.
concentration point of TOC RANGE 2 in System Range Data
menu by default.
BioTector to change its range automatically from range 3 to
range 1.
concentration point of TOC RANGE 1 in System Range Data
menu by default.
injected into the reactor at range 3.
injected into the reactor at range 3.
the reactor for the BASE OXIDATION phase at range 3.
which can be programmed independently for range 3.
in System Program 1 menu, depending on the difference
between the two settings, the reaction time at range 3 will be
either longer or shorter than the reaction time displayed in
System Program 1 menu.
into the reactor for the TOC SPARGE phase at range 3.
Page 117