Immunity Test
Surge EN/IEC
61000-4-5
(only for devices with battery
charger power supply
-7000MHB & 7100MHB)
Power Frequency
50/60Hz Magnetic
Field EN/IEC
61000-4-8
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines EN/
IEC 61000-4-11
(only for devices with
battery charger power
supply -7000MHB &
7100MHB)
NOTE : U
is the a.c. mains voltage prior to application of the test level.
T
EN/IEC 60601
Compliance
Test Level
Level
±1 kV line(s)
to line(s)±2 kV
to line(s)±2 kV
line(s) to earth
line(s) to earth
3A/m
<5% U
<5% U
T
(>95% dip in U
)
(>95% dip in U
T
for 0.5 cycle.
for 0.5 cycle.
40% U
(60% dip
40% U
T
in U
) for 5
in U
T
cycles.
cycles.
70% U
(30% dip
70% U
T
in U
) for 25
in U
T
cycles
cycles
5% U
(>95% dip
5% U
T
in U
) for 5 sec.
in U
T
Electromagnetic
Environment – Guidance
±1 kV line(s)
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency magnetic
3A/m
fields should be that of a
typical commercial or hospital
environment.
T
Electrical power quality
)
should be that of a typical
T
commercial or hospital
environment. During
(60% dip
T
charging, if the the user of the
) for 5
GoSafe (7000MHB/7100MHB)
T
requires continued operation
during power mains
interruptions, it is
(30% dip
T
recommended that the
) for 25
T
GoSafe device power adapter
is powered from an
uninterrupted power supply
(>95% dip
T
or a battery.
) for 5 sec.
T