Vega VEGAPULS 68 Operating Instructions Manual

4...20 ma/hart two-wire, sensor for continuous level measurement of bulk solids under simple process conditions, suitable for steel industry applications
Hide thumbs Also See for VEGAPULS 68:

Advertisement

Operating Instructions

VEGAPULS 68
4 ... 20 mA/HART two-wire
Radar

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VEGAPULS 68 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Vega VEGAPULS 68

  • Page 1: Operating Instructions

    Operating Instructions VEGAPULS 68 4 … 20 mA/HART two-wire Radar...
  • Page 2 Menu schematic ......45 Saving the parameter adjustment data ... . 47 Radar VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 3 34296 - Protective cover 31088 - Flanges according to DIN-EN-ASME-JIS 30176 - Electronics module VEGAPULS series 60 31381 - Antenna impedance cone VEGAPULS 62 and 68 VEGAPULS 68 • 4 … 20 mA/HART two-wire VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 4: Function

    Modifications not expressly approved by VEGA will lead to expiry of the operating licence according to FCC. VEGAPULS 68 is in conformity with part 15 of the FCC regulations. Take note of the respective operating regulations: VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 5: Product Description 3.1 Configuration

    Please help us fulfil this obligation by observing the environmental instructions in this manual: Components Chapter "Packaging, transport and storage" Chapter "Disposal" Fig. 1: VEGAPULS 68 - horn antenna and swivelling holder Housing cover with integrated PLICSCOM (optional) Housing with electronics Swivelling holder with flange Horn antenna VEGAPULS 68 consists of the following components: VEGAPULS 68 •...
  • Page 6: Principle Of Operation

    DIN EN 24180. A version of VEGAPULS 68 is available for each area of application: The packaging of standard instruments consists of environment- The version with horn antenna is particularly suitable for friendly, recyclable cardboard.
  • Page 7: Mounting 4.1 General Instructions

    The reference plane for the measuring range of the sensors is the lower edge of the flange or the seal surface of the thread. VEGAPULS 68 • 4 … 20 mA/HART two-wire VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 8: Mounting Preparations - Horn Antenna

    This information applies only to special versions! the polarisation plane is marked on the instrument. VEGAPULS 68 is also available in versions where the antenna has a bigger diameter than the process fitting (thread, flange). The antenna must therefore be disconnected from the process fitting before mounting.
  • Page 9: Mounting Preparations - Parabolic Antenna

    10 Tighten locknut (2) with SW 36, torque max. 40 Nm. Note: Take note for VEGAPULS 68 with rinsing air connection that the holes in the antenna and in the process fitting correspond. This ensures a sufficient air flow (the air is led through the holes to the feed system. A rinsing of the parabolic antenna in total is not intended).
  • Page 10 The following description gives an overview on the determination of the necessary angle of inclination. VEGAPULS 68 • 4 … 20 mA/HART two-wire VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 11 Proceed as follows to adjust the angle of inclination with the swivelling holder: Loosen terminal screw of the swivelling holder with a fork spanner Fig. 9: Proposal for installation after orientation VEGAPULS 68 SW 13 Direct the sensor, check angle of inclination The angle of inclination depends on the vessel dimensions.
  • Page 12 ‰ Fig. 12: Distance between antenna and socket Fig. 10: Inflowing medium If the medium has good reflective properties, VEGAPULS 68 can also Socket The socket piece should be dimensioned in such a way that the be mounted on a longer socket piece. Recommended values for antenna end protrudes slightly out of the socket.
  • Page 13 Fig. 15: Orientation of VEGAPULS 68 for emptying in the silo center Vessel installations Silo installations such as e.g. ladders, level switches, struts, and also structured vessel walls, can cause false echoes that get superimposed on the useful echo.
  • Page 14 Tip: sensors is not possible. In case of extreme dust in the antenna system, VEGAPULS 68 is available with rinsing connection e.g. for air. The air is distributed via channels in the antenna system and keeps it virtually free of dust.
  • Page 15: Connecting To Power Supply 5.1 Preparing The Connection

    Provide a reliable separation between the supply circuit and the mains circuits according to DIN VDE 0106 part 101. The VEGA power supply units VEGATRENN 149A Ex, VEGASTAB 690 as well as all 5.2 Connection procedure VEGAMETs meet this requirement.
  • Page 16: Wiring Plan, Single Chamber Housing

    Plug connector for VEGACONNECT (I²C interface) Spring-loaded terminals for connection of the external indication VEGADIS Ground terminal for connection of the cable screen Spring-loaded terminals for voltage supply VEGAPULS 68 • 4 … 20 mA/HART two-wire VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 17: Wiring Plan, Double Chamber Housing

    Fig. 26: Connection compartment, double chamber housing Plug connector for VEGACONNECT (I²C interface) Ground terminal for connection of the cable screen Spring-loaded terminals for voltage supply VEGAPULS 68 • 4 … 20 mA/HART two-wire VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 18: Wiring Plan Double Chamber Housing Ex D

    Fig. 30: Connection compartment double chamber housing Ex d Spring-loaded terminals for power supply and cable screen Ground terminal for connection of the cable screen VEGAPULS 68 • 4 … 20 mA/HART two-wire VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 19: Wiring Plan - Version Ip 66/Ip 68, 1 Bar

    (you can choose any one of four different Switch on phase After connecting VEGAPULS 68 to power supply or after a voltage positions - each displaced by 90°) recurrence, the instrument carries out a self-check for approx. 30...
  • Page 20: Insert Indicating And Adjustment Module

    Approx. 10 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not confirmed with [OK] will not be saved. VEGAPULS 68 • 4 … 20 mA/HART two-wire VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 21: Setup Procedure

    Carrying out max. ad- Proceed as follows: justment Parameter adjustment As VEGAPULS 68 is a distance measuring instrument, the distance Max. adjustment from the sensor to the product surface is measured. To have the real 100.00 % product level displayed, an allocation of the measured distance to the percentage height must be made.
  • Page 22 They represent the correlation between the level percentage level change VEGAPULS 68 for applications in which the level changes very and vessel volume. By activating the appropriate curve, the volume quickly. For this reason, select the function "Quick level change > 1 m/ percentage of the vessel is displayed correctly.
  • Page 23: Menu Schematic

    HART mode no reset Special parameters are parameters which are set customer-specifically on Sensor-specific basic adjustment. the service level with the adjustment software PACTware. VEGAPULS 68 • 4 … 20 mA/HART two-wire VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 24 Last change using PC Sensor characteristics 21. January 2009 Basic adjustment Software version Display now? Display ▶ Serial number 3.50 21. January 2009 Diagnostics 12345678 Service Info VEGAPULS 68 • 4 … 20 mA/HART two-wire VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 25: Saving The Parameter Adjustment Data

    They are hence 7.1 Connect the PC via VEGACONNECT available for multiple use or service purposes. If VEGAPULS 68 is equipped with an indicating and adjustment Internal connection via module, the most important data can be read out of the sensor into I²C interface...
  • Page 26: Parameter Adjustment With Pactware

    7 Setup with PACTware and other adjustment programs 7 Setup with PACTware and other adjustment programs Necessary components: Note: Keep in mind that for setup of VEGAPULS 68, DTM-Collection in the VEGAPULS 68 actual version must be used. PC with PACTware and suitable VEGA DTM...
  • Page 27: Maintenance And Fault Rectification 8.1 Maintenance, Cleaning

    E041, E042, E043 24 hour service hotline However, should these measures not be successful, call the VEGA Hardware error, electronics defective service hotline in urgent cases under the phone no. +49 1805 858550.
  • Page 28: Software Update

    Connect the sensor to power supply and provide connection from PC to the instrument via VEGACONNECT. Start PACTware and provide connection to the sensor, e.g. via the VEGA project assistant. Close the parameter window of the sensor, as far as open.
  • Page 29: Dismounting 9.1 Dismounting Steps

    Process fitting swivelling holder with 6 … 17.2 kg (13.23 … 37.92 lbs) flange, depending on flange size and housing Output variable Output signal 4 … 20 mA/HART VEGAPULS 68 • 4 … 20 mA/HART two-wire VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 30 860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig) Other reference conditions Reflector ideal reflector, e.g. metal plate 2 x 2 m Fig. 38: Deviation VEGAPULS 68 with horn antenna False reflections Biggest false echo, 20 dB smaller than the useful echo...
  • Page 31 Relating to the nominal measuring range, in the temperature range -40 … +80 °C . Tested according to the regulations of German Lloyd, GL directive 2. VEGAPULS 68 • 4 … 20 mA/HART two-wire VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 32 25 mm (0.984 in) with 25 °C (77 °F) Depending on the version M12 x 1, according to DIN 43650, Harting, 7/ 8" FF. VEGAPULS 68 • 4 … 20 mA/HART two-wire VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 33: Technical Data

    For these instruments, the corresponding approval documents have to be taken into account. 1000 These are part of the delivery or can be downloaded under www.vega.com via "VEGA Tools" and "serial number search" as well as via "Downloads" and "Approvals".
  • Page 34: Dimensions

    Aluminium housing Aluminium double chamber housing Stainless steel housing, electropolished Stainless steel housing - precision casting Stainless steel double chamber housing - precision casting VEGAPULS 68 • 4 … 20 mA/HART two-wire VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 35 Fig. 44: VEGAPULS 68 - horn antenna in threaded version Standard with temperature adapter Fig. 45: VEGAPULS 68 - horn antenna in threaded version with purging air connection Standard with temperature adapter Purging air connection G⅛ A for mounting of a suitable adapter Reflux valve - unmounted (with non-Ex optionally available, with Ex in the scope of delivery), for tube diameters...
  • Page 36 11.00" 1.00" 9.50" 8xø0.88" 3.74" 16.93" Fig. 46: VEGAPULS 68 - horn antenna in threaded version with antenna extension Standard Fig. 47: VEGAPULS 68 - horn antenna in flange version with temperature adapter Standard with temperature adapter An antenna extension causes a reduction of sensitivity at close range in dependence on product properties.
  • Page 37 228,6 11,5 190,5 8x ø 19,1 4" Fig. 48: VEGAPULS 68 - horn antenna in flange version with purging air connection Standard Fig. 49: VEGAPULS 68 - horn antenna and swivelling holder with temperature adapter Standard Purging air connection G⅛ A for mounting of a suitable adapter with temperature adapter Reflux valve - unmounted (with non-Ex optionally available, with Ex in the scope of delivery), for tube diameters...
  • Page 38 8x ø 19,1 4" Fig. 50: VEGAPULS 68 - horn antenna, swivelling holder and purging air connection Purging air connection G⅛ A for mounting of a suitable adapter Reflux valve - unmounted (with non-Ex optionally available, with Ex in the scope of delivery), for tube diameters 6 mm VEGAPULS 68 •...
  • Page 39 8x ø 19,1 4" 9" 0.45" 7.5" 8x ø 0.75" Fig. 52: VEGAPULS 68 - parabolic antenna in flange version Standard with temperature adapter Fig. 53: VEGAPULS 68 - parabolic antenna and swivelling holder Standard with temperature adapter VEGAPULS 68 • 4 … 20 mA/HART two-wire...
  • Page 40 11,5 190,5 8x ø 19,1 4" Fig. 54: VEGAPULS 68 - parabolic antenna and swivelling holder with purging air connection Standard with temperature adapter Purging air connection G⅛ A for mounting of a suitable adapter Reflux valve - unmounted (with non-Ex optionally available, with Ex in the scope of delivery), for tube diameters 6 mm VEGAPULS 68 •...
  • Page 41 © VEGA Grieshaber KG, Schiltach/Germany 2009 Subject to change without prior notice 29261-EN-090305 VEGAPULS 68 • 4 … 20 mA/HART two-wire...