Air flow
functionality
16

Air Flow functionality

Air Flow functionality
(Available using 3M's Air Probe on QUESTempº 36 model.)
Air flow is measured in meters per second over a range of 0 to 20m/s in 0.1m/s
increments. The sensor should be placed or held perpendicular in the air stream.
Unlike many anemometers, the omni-directional sensor does not require rotating to
find the maximum reading. Be careful not to block the air flow with your body
during measurements. The sensor's measuring tip is fragile; be cautious if
measuring in ducts.
The Air Probe may be either hand held or mounted behind the QUESTemp° 36
using the mounting bracket hooked to the sensor bar beneath the center bulb
sensor. (See Figure 1-21 below.)
A green lamp indicator in the Air Probe indicates that it is turned on and the battery
is good. If the green indicator turns off while the switch is in the On position,
replace or recharge the battery.
To connect, follow steps 1-3 below.
1. Clip front of bracket around lip of
sensor bar.
2. Tighten bracket to sensor bar with
3. Attach air probe to bracket
thumb screw.
with thumb screen
Bracket
QTº36 sensor bar
Air Probe
Figure 1-21: Air flow configuration