Table 4- 10: Diagnostic error messages (Cont.)
Module
RSA6100A Series Service Manual Serial Numbers B020000 and Above
Test
Error Message
The - - 8 Volt supply is under voltage.
The - - 8 Volt supply is under voltage.
The - - 15 Volt supply is over voltage.
The - - 15 Volt supply is under voltage.
The - - 5 Volt supply is over voltage.
The - - 5 Volt supply is under voltage.
Maintenance
Recommended Action
Replace the Power Conversion Board.
If the problem persists, replace the RF
Interface Board. If the problem
persists, replace the power supply. If
the problem persists, send the instru-
ment to a Tektronix Service Center for
repair.
Replace the Power Conversion Board.
If the problem persists, replace the RF
Interface Board. If the problem
persists, replace the power supply. If
the problem persists, send the instru-
ment to a Tektronix Service Center for
repair.
Replace the Power Conversion Board.
If the problem persists, replace the RF
Interface Board. If the problem
persists, replace the power supply. If
the problem persists, send the instru-
ment to a Tektronix Service Center for
repair.
Replace the Power Conversion Board.
If the problem persists, replace the RF
Interface Board. If the problem
persists, replace the power supply. If
the problem persists, send the instru-
ment to a Tektronix Service Center for
repair.
Replace the Power Conversion Board.
If the problem persists, replace the RF
Interface Board. If the problem
persists, replace the power supply. If
the problem persists, send the instru-
ment to a Tektronix Service Center for
repair.
Replace the Power Conversion Board.
If the problem persists, replace the RF
Interface Board. If the problem
persists, replace the power supply. If
the problem persists, send the instru-
ment to a Tektronix Service Center for
repair.