Unit ID Setup
– Associate unit with an ID number
Unit #
000
Keypad:
Unwrapped
Wrapped
Rubber and Plastics
Stop
Set cycle counter
– Adjust the recorded number of cycles ran
Cycle Number
000000
Keypad:
Unwrapped
Wrapped
Rubber and Plastics
Stop
Conductivity Setup
CD= x.xuS/NNN/y.yppm
L=LL.L H=HH.H G=G.GG
If Statim has a Float Switch and Water Quality Sensor display below will appear.
CD= x.xuS/NNN/y.yppm
FLOAT H=HH.H G=G.GG
Screen Representation
x.x
NNN
y.y
LL.L
FLOAT
HH.H
G.GG
Note: Distilled water readings should be between Low and High thresholds.
Go to Water Conductivity Circuit Calibration to adjust readings.
Keypad:
Unwrapped
Wrapped
Rubber and Plastics
Stop
Note: To perform Water Conductivity Circuit Calibration see page 18.
:
Decrease current field (the flashing value on the display)
Increase current field (the flashing value on the display)
Select next digit
Save & exit menu to normal mode of operation
Decrease current digit
Increase current digit
Select next digit
Save & exit menu to normal mode of operation
– To display detected water quality
Water conductivity in uS (micro-Siemens)
Water conductivity in ADC (Analog to Digital converter) counts (0...255)
Water quality in ppm (parts per million)
Lower value threshold (No water threshold)
Values lower than this trigger "No water refill reservoir" error
Indicates the reservoir has both a Float Switch & Water Quality Sensor
High value threshold (Bad water threshold)
Values larger than this trigger "Bad water quality" error
Water conductivity circuit gain
Increase current field
Decrease current field
Move to next field
Exit menu to normal mode of operation
(Used with printer)
default 0.3uS
default 10uS
default 1.00
9