IDA-5
Users Manual

Storage

Remove all water from the Product before storage, particularly if temperatures can fall
below 5 °C (41 °F). Do not pressurize the inlet ports. It is safest to use a medical suction
pump to drain the measuring channels and use the Cycle Valves from the Calibration
menu (follow on-screen instructions).

Shipping

Remove all liquid from the Product before shipping. To prevent liquid from entering the
ports, put the Product in a large plastic bag. Put the bagged Product into its shipping
carton. If this is not available, make sure there is shock protection with a minimum of
5 cm compressible cushioning inside the carton (e.g. 60 cms x 60 cms x 60 cms).

Specifications

General Specifications

Operating Voltage Range ........... 100 to 240 V ac
Supply Frequency ....................... 50/60 Hz
Supply Power ............................. <50 VA
Fuses ........................................... 20 mm T1.6 A H 250 V x 2
Size ............................................. 30 H x 20 W x 20 D (cm) 12 x 8 x 8 (in.)
Weight ........................................ 5 kg (approx) (11 lbs.)
Altitude ....................................... 0 to 3000 m (10000 ft)
Temperature

Performance Specifications

Flow Rate Measurement
30
Operating ....................... 15 °C to 30 °C (59 °F to 86 °F)
Storage ........................... -20 °C to +40 °C (-4 °F to +104 °F) when drained of all
Humidity ........................ 10 % to 90 % non-condensing
Method ........................... Flow is calculated by measuring volume over time
Range ............................. 0.1 to 1500 ml/h (2500 ml/h is shown)
Accuracy ....................... 1 % of reading ±1 LSD for flows of 16 to 200 ml/h for
Max Test Duration ........ 100 hours
liquid
volumes over 20 ml, otherwise 2 % of reading ±1 LSD
for volumes over 10 ml under laboratory conditions.
Degassed water at 15 to 30 °C (59 to °86 F) is
recommended for long tests. Accuracy for >16 ml/hr
and >1500 ml/hr not specified.