Replacing the Filter Element in the Probe,
Replacing the
Filter Element
26
Step
Action
1
Turn the T-handle 1 of the filter
removal device 1-3 in counter-
clockwise direction.
This pulls the filter element 8 via the
detaching disk 3 out of the casing
10.
2
Turn bridge 2 until it can be pulled off
from the bridge holding device 9
through the elongated holes.
3
Pull out filter element 8 with bridge 2
and detaching disk 3.
4
Turn detaching disk 3 until it can be
pulled off from the hexagon screws 5
via the elongated holes.
Never loosen or tighten the hexagon screws 5. They have been
adjusted at the factory so that the detaching disk 3 can be
easily moved.