Multi-Fid" Screen - ABB ACF-NT Operator's Manual

Multicomponent analyzer systems for emission and process monitoring
Hide thumbs Also See for ACF-NT:
Table of Contents

Advertisement

"Multi-FID" Screen

Figure 7
"Multi-FID" Screen
Indication
18
The "Multi-FID" screen displays the values of the VOC analyzer temperature and
pressure regulators.
Parameter
T-Co.D
Detector temperature (on "ASP Module" screen)
T-Co.E
Sample gas port temperature
Flame
Flame temperature
A.Pres
Combustion air pressure
H2
Combustion gas pressure
Input
Instrument air pressure at combustion chamber
inlet
Output
Instrument air pressure at combustion chamber
outlet
Operator's Manual ACF-NT, ACF-NT V0309
Default
180 °C
180 °C
260–300 °C
725 hPa
1100 hPa
700 hPa
600 hPa
42/23-572 EN Rev. 5