VOC Analyzer Calibration
Zero Calibration
Span Calibration
Maintenance Interval
Calibration Control
Calibration Data
42/23-572 EN Rev. 5
For the zero calibration either nitrogen from a cylinder or the zero gas for the
oxygen analyzer is used. In the latter case a T-piece must be used to connect the
test gas cylinder with the test gas port and the FID zero gas port of the analyzer
system.
For the span calibration n-Propane in nitrogen from a test gas cylinder is used.
According to the officially established maintenance interval zero and span calibra-
tion have to be performed every 14 days.
The VOC analyzer calibration is performed as externally controlled automatic
calibration.
The FTIR controller triggers the start of the automatic calibration.
Enable
Test gas concentration
Single zero calibration
Single span calibration
Common zero and span calibration Always
Operator's Manual ACF-NT, ACF-NT V0309
Off
Zero
0
Span
Bottle concentration
Never
Never
37