Chapter 3
Safety Notes
Safety Labels
Affixed to the
Analyzer System
Harmful Gases
Cleaning Hints
Cleaning Hints
42/23-572 EN Rev. 5
Maintenance
CAUTION!
Only persons familiar with the maintenance of comparable analyzer systems
and certified as being capable of such work should work on the system.
CAUTION!
Observe the safety labels affixed to the analyzer system:
Consult Documentation!
Hot Surface! (Temperature > 60 °C)
Risk of Electric Shock!
WARNING!
Some of the gases measured with the analyzer system are harmful to health.
Therefore, the sample gas must not escape from the gas path during normal
operation and maintenance works.
A seal integrity check of the analyzer system has to be performed at regular
intervals.
The diluted exhaust gas must be drained out of the installation room of the
analyzer cabinet.
Never use water or any solvents to clean parts inside the analyzer cabinet.
The analyzer system shall always operate with cabinet door closed. If dust
invades the cabinet it shall be removed using a broom and a vacuum cleaner.
The outside of the analyzer cabinet can be cleaned with a wet towel and mild
cleaning agents. Pay attention that no droplets invade the cabinet.
Operator's Manual ACF-NT, ACF-NT V0309
21