Siemens Maxum II Reference Manual page 89

Hide thumbs Also See for Maxum II:
Table of Contents

Advertisement

Display
panel to
TIB cable
Figure 2-47
TIB Color Touchscreen Door Exploded View
9. Remove the TIB protective cover by removing the 4 screws. (A in exploded view) This
allows access to two of the 6 mounting studs for the LCD assembly. (E in exploded view)
10.Disconnect the three connectors of the display cable assembly; the backlight connector,
the TIB cable connector, and the LCD connector. (B in exploded view)
Maxum II Reference Manual
Manual, 7/2017, 2000596-001
Electronic Compartment Component Descriptions and Maintenance Procedures
B
LCD
from
SIB3
Intrinsic safety
ground wires
E
B
E
Hole for
wire-tie
E
A
C
E
Backlight
Connector
E
Hole for
wire-tie
2.10 Color Touchscreen
D
E
Wire-tie
anchors,
insertion
points
D
ESD
ground
connection
points
89