Omega DP41-S User Manual page 95

High performance strain gage indicator
Hide thumbs Also See for DP41-S:
Table of Contents

Advertisement

2. Now you are in the run mode and the transducer is connected to
the meter. Simulate two loads on the transducer and record the
meter displays as "INPUT1" and "INPUT2" and record what
each of these readings ("READ 1" and "READ 2") should
display on a piece of scrap paper or in the spaces provided below.
These numbers will be the INPUT values entered in the
"IN.SC.OF" scaling. Normally the two points that are calibrated
are the no load condition and 3/4 full scale. However, you may
use any two measurements.
"INPUT1"
"READ 1"
Assume that you simulate no displacement and the display indi-
cates "-55.6", and when you simulate 4.0000 inches the display
indicates "16136.6".
The following data will now be entered into the two point scaling
"IN.SC.OF" method.
"INPUT1":
-0055.6
"INPUT2":
16136.6
Ignore the decimal point position for the READ values. Just note
how many zero's (or places to the right of the decimal) you need.
In this example there were 4 decimal places needed and there are 4
zeros after the "4" in the "READ 2" value (ignore the decimal
point).
You must alter the "RDG SC" factor as follows. Note that the
decimal point above in the "READ 1" and "READ 2" values is not
in the correct position. You need to move it 3 places to the left to
make it correct.
By looking at Table 8-6, you can see that a value of 0.00100 must
be entered for a new "RDG SC" value.
"INPUT2"
"READ 2"
"READ 1":
00000.0
"READ 2"
04000.0
8-51
8
Scaling to
Display
Engineering
Units