Omega DP41-S User Manual page 56

High performance strain gage indicator
Hide thumbs Also See for DP41-S:
Table of Contents

Advertisement

8
Scaling to
Display
Engineering
Units
MENU
SUBMENU 1 SUBMENU 2 DESCRIPTION
IN CNF
INP.1=0
INP.2=0
INP.3=0
INP.4=0
INP.5=
INP.6=0
INP.7=1
IN.SC.OF
DEC PT
FFFFF.F
CNT BY
2. Now you are in the run mode and the load cell is connected to
the meter. Simulate two loads on the transducer and record the
meters display as "INPUT1" and "INPUT2" and record what
each of these readings ("READ 1" and "READ 2") should display
in the spaces provided below. These numbers will be the INPUT
values entered in the "IN.SC.OF" scaling. Normally the two
points that are calibrated are the no load condition and 3/4 full
scale. However, you may use any two measurements.
"INPUT1"
"READ 1"
Assume that you simulate no load the display indicates "–55.6",
and when you simulate 4.0000 pounds the display indicates
"16136.6".
60 Hz ac power
Slow reading
(S2A jumper omitted)
Unipolar input
(S2B jumper omitted)
Std, for current inputs
Disregard not used
Disables IN.SC.OF
Ratiometric input
Skip past
By shifting the decimal
point, it makes the meter
10 times more sensitive
(desired decimal point
will be entered later)
Press the 'RESET' button
two times to return to
the run mode
"INPUT2"
"READ 2"
8-12