Omega DP41-S User Manual page 104

High performance strain gage indicator
Hide thumbs Also See for DP41-S:
Table of Contents

Advertisement

9
Meter Function Menus
MENU
BUTTON
MAIN MENU
"RDG.CNF":
* The 'MIN' button allows you to sequence through RDG.1, RDG.2, RDG.3,
RDG.4, RDG.5, RDG.6, and RDG.7.
The 'MAX' button allows you to select the "0" or "1" state for each "RDG"
condition.
The 'MENU' button stores the selected values for all "RDG.CNF"
condition(s) changed and advances the meter to "RD SC". Do not press the
'MENU' button after each change within the submenu or the meter will
advance to the next menu item.
Every underlined "0" or "1" state is the factory preset value.
9-8
MIN/MAX/MENU *
BUTTON
SUB MENU
"RDG.1=0"
"RDG.1=1"
"RDG.2=0"
"RDG.2=1"
"RDG.3=0"
"RDG.3=1"
"RDG.4=0"
"RDG.4=1"
"RDG.5=0"
"RDG.5=1"
"RDG.5=2"
"RDG.6=0"
"RDG.6=1"
"RDG.7=0"
"RDG.7=1"
DESCRIPTION
READING CONFIGURATION
Reading Scale & Offset:
Direct Format
2-Coordinate format
Decimal point effect:
Active
Independent
Display Brightness:
Normal
50% of Normal
Leading Zeros on Display:
Displayed
Not Displayed
Unit of Temperature:
NOT USED WITH
STRAIN GAGE
Reading Scale & Offset Functions:
Disables Reading Scale and Offset
(RDG SC=1.0, RDG OF=0.0)
Enables Reading Scale and Offset
(Uses scale and offset values entered
by user)
Resetting Mode:
Grounding P2-5 causes
Hard Reset (RESET 2)
Grounding P2-5 causes Peak/Valley
(HI/LO) Reset