5. Troubleshooting

The following is a list of problems that may occur while using the BiliTx system. For additional
information, refer to the BiliTx Service Manual or contact Respironics Customer Service at 1-800-345-
6443 or 1-724-387-4000.
WARNING
If therapy is interrupted for one hour or longer, resume therapy as soon as possible.
Problem
Therapy On/Off button is not
green
Therapy On/Off button is not
flashing green
Light is not being emitted from
fiber optic panel
Yellow LED is flashing
Loss of power or light source
failure
Light is not being emitted from
the illuminator device when
connected to the circuit support
adapter and the yellow LED is
flashing
Device will not turn off when
the Therapy On/Off button is
pressed
BiliTx™
professional manual
Check to make sure power cord is properly attached and
plugged into an active electrical outlet.
Ensure the device is turned on.
If the power cord is properly attached and plugged into an
active electric outlet and the Therapy On/Off button is not
flashing green, but the device will turn on, continue to use the
device for therapy.
Check to make sure panel is securely locked into Illuminator
device.
Ensure the panel or circuit support adapter is securely locked
into place. If light continues to flash, contact Respironics.
Contact Respironics.
Check to make sure the Illuminator device is securely attached
to the circuit support adapter accessory.
Press and hold the Therapy On/Off button for 3 seconds.
Reason/Action