4. Connect the circuit support adapter to the clamp on the circuit
support arm. Tighten in place by turning the handle on the
clamp.
5. Attach the device power cord to the circuit support arm.
6. Position the device over the baby ensuring that it is no closer
than 6 inches from the baby. The intensity of light can be
adjusted by positioning the light source closer or farther away
from the baby. See the light intensity chart in the Specifications
chapter of this manual for the approximate light intensity at
specific distances from the device.
7. Cover the baby's eyes with the appropriate eye protection. Refer
to the instructions included with your eye protection.
8. Plug the Illuminator device into an electrical outlet. The power
button will flash green.
9. Press the Therapy On/Off button to turn the Illuminator device
on. The Therapy button will illuminate green.
10. To turn off the device when therapy is complete, press and hold
the Therapy On/Off button for 3 seconds.
BiliTx Professional Manual
Connecting the circuit
support adapter to the
circuit support arm
WARNING
The power cord must be
attached to the circuit support
arm to avoid entanglement.
WARNING
Do NOT position the device
closer than 6 inches from the
baby.
WARNING
Verify that the Illuminator unit is
securely attached to the circuit
support arm and the mounting
brace is securely attached to the
IV pole before placing the baby
under the device.
WARNING
Place the IV pole supporting
the unit so that it will not be
accidentally disturbed during
overhead therapy.
CAUTION
If the power cord or wires need
repair or replacement, do not
connect the device.