Serial Interface Parameters

The serial interface menu is accessed by pressing "UP" from the operating pa-
rameters menu.
The serial RS-232 interface parameters menu is indicated by,
Serial RS-232 interface parameters menu
SErIAL
The serial interface parameters menu contains parameters which determine the
operation of the serial interface. The parameters in the menu are—baud rate,
sample period, duplex mode, and linefeed.

Baud Rate

The baud rate is the first parameter in the menu. The baud rate setting deter-
mines the serial communications transmission rate.
The baud rate parameter is indicated by,
Serial baud rate parameter
bAUd
Press "SET" to choose to set the baud rate. The current baud rate value will
then be displayed.
1200 b
Current baud rate
The baud rate of the bath serial communications may be programmed to 300,
600, 1200, or 2400 baud. Use "UP" or "DOWN" to change the baud rate value.
2400 b
New baud rate
Press "SET" to set the baud rate to the new value or "EXIT" to abort the opera-
tion and skip to the next parameter in the menu.

Sample Period

The sample period is the next parameter in the serial interface parameter menu.
The sample period is the time period in seconds between temperature measure-
ments transmitted from the serial interface. If the sample rate is set to 5, the
bath transmits the current measurement over the serial interface approximately
every five seconds. The automatic sampling is disabled with a sample period of
0. The sample period is indicated by,
Serial sample period parameter
SAnPLE
Press "SET" to choose to set the sample period. The current sample period
value will be displayed.