Serial Interface Parameters; Baud Rate - Fluke 7012 User Manual

Calibration bath
Table of Contents

Advertisement

CtorSt
Press "SET" to access the parameter setting. Normally the cutout is set for
manual mode.
Cto=rSt
To change to automatic reset mode press "UP" and then "SET".
Cto=Auto
9.12

Serial Interface Parameters

The 7012 bath may optionally be fitted with an RS-232 interface. The serial
RS-232 interface parameters menu is indicated by,
SErIAL
The serial interface parameters menu contains parameters which determine the
operation of the serial interface. These controls only apply to baths fitted with
the serial interface. The parameters in the menu are — BAUD rate, sample pe-
riod, duplex mode, and linefeed.
9.12.1

Baud Rate

The baud rate is the first parameter in the menu. The baud rate setting deter-
mines the serial communications transmission rate.
The baud rate parameter is indicated by,
BAUd
Press "SET" to choose to set the baud rate. The current baud rate value will
then be displayed.
1200 b
The baud rate of the bath serial communications may be programmed to
300,600,1200, or 2400 baud. Use "UP" or "DOWN" to change the baud rate
value.
2400 b
Press "SET" to set the baud rate to the new value or "EXIT" to abort the opera-
tion and skip to the next parameter in the menu.
Cutout reset mode parameter
Cutout set for manual reset
Cutout set for automatic reset
Serial RS-232 interface parameters menu
Serial baud rate parameter
Current baud rate
New baud rate
9 Controller Operation
Serial Interface Parameters
47