7341 Calibration Bath
User's Guide
1.2
1.2.1
2
Symbol
Description
C-TIC Australian EMC mark
The European Waste Electrical and Electronic Equipment (WEEE) Directive
(2002/96/EC) mark.

Safety Information

Use this instrument only as specified in this manual. Otherwise, the protection
provided by the instrument may be impaired. Refer to the safety information in
Warnings and Cautions sections.
The following definitions apply to the terms "Warning" and "Caution".
• "WARNING" identifies conditions and actions that may pose hazards to
the user.
• "CAUTION" identifies conditions and actions that may damage the in-
strument being used.

WARNINGS

To avoid possible electrical shock or personal injury, follow these guidelines.
GENERAL
• DO NOT use the instrument for any application other than calibration
work. The instrument was designed for temperature calibration. Any other
use of the instrument may cause unknown hazards to the user.
• DO NOT use the instrument in environments other than those listed in
the user's guide.
• DO NOT overfill the bath. Overflowing extremely cold or hot fluid may
be harmful to the operator. See Section 5.3, Bath Preparation and Filling,
for specific instructions.
• Follow all safety guidelines listed in the user's manual.
• Calibration Equipment should only be used by Trained Personnel.
• If this equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
• Before initial use, or after transport, or after storage in humid or semi-hu-
mid environments, or anytime the instrument has not been energized for
more than 10 days, the instrument needs to be energized for a "dry-out"
period of 2 hours before it can be assumed to meet all of the safety re-
quirements of the IEC 1010-1. If the product is wet or has been in a wet
environment, take necessary measures to remove moisture prior to apply-
ing power such as storage in a low humidity temperature chamber operat-
ing at 50°C for 4 hours or more.