7341 Calibration Bath
User's Guide
9.3.2
9.3.3
40
S
Accept selected set-point memory

Set-point Value

After selecting the set-point memory and pressing "SET", the set-point value
may be adjusted in increments of 0.01° (C or F). The set-point value is dis-
played with the units, C or F, at the left.
Set-point 4 value in °C
C
40.00
If the set-point value does not need to be changed, press "EXIT" to resume dis-
playing the bath temperature. Press "UP" or "DOWN" to adjust the set-point
value.
U
Increment display
C
42.50
New set-point value
When the desired set-point value is reached, press "SET" to accept the new
value and access the set-point vernier. If "EXIT" is pressed, any changes made
to the set-point are ignored.
S
Accept new set-point value

Set-point Vernier

The user may want to adjust the set-point slightly to achieve a more precise
bath temperature. The set-point vernier allows one to adjust the temperature be-
low or above the set-point by a small amount with very high resolution. Each of
the 8 stored set-points has an associated vernier setting. The set-point vernier
can be set in increments of 0.00018°C. The vernier is accessed from the
set-point by pressing "SET". The vernier setting is displayed as a 6 digit num-
ber with five digits after the decimal point. This is a temperature offset in de-
grees of the selected units, C or F.
Current vernier value in °C
0.00000
To adjust the vernier, press "UP" or "DOWN". Unlike most functions the ver-
nier setting has immediate effect as the vernier is adjusted. "SET" need not be
pressed. This allows the bath temperature to be continually adjusted as it is
displayed.
U
Increment display
New vernier setting
0.00090