Quick Start

CAUTION: Read Section 6 entitled BATH USE before placing the bath in
service.
This chapter gives a brief summary of the steps required to set up and operate
the bath. This should be used as a general overview and reference and not as a
substitute for the remainder of the manual. Please read Section 5, Installation,
through Section 8, General Operation, carefully before operating the bath.
4.1

Unpacking

Unpack the bath carefully and inspect it for any damage that may have oc-
curred during shipment. If there is shipping damage, notify the carrier immedi-
ately. Verify that all components are present:
• 7341 Bath
• Access Hole Cover
• Fluid Expansion Reservoir with Cover
• User's Guide
• RS-232 Cable
• Report of Test
• Drain Valve
• 9930 Interface-it Software and User's Guide
• Tipping Prevention Bracket, including mounting hardware for wood or
concrete floor
• 2019-DCB Kit, Liquid-in-Glass (LIG) Adapter and Carousel (optional)
• 2069 Scope, Liquid-in-Glass (LIG) with mounts (optional)
If you are missing any item, please call an Authorized Service Center. To locate
an Authorized Service Center near you see Section 1.3 on page 5.
4.2

Set Up

WARNING: The instrument is equipped with a soft cutout (user settable
firmware) and a hard cutout (set at the factory). Check the flash point,
boiling point, or other fluid characteristic applicable to the circumstances
of the instrument operation. Ensure that the soft cutout is adjusted to the
fluid characteristics of the application. As a guideline, the soft cutout
should be set 10°C to 15°C below the flash point of the bath fluid. See
Section 8.1, Heat Transfer Fluid, for specific information on bath fluids
and Section 9.10, Cutout.