Taking measurements in 1x, Manual and
«Manual» Mode
Select «Manual» Mode if auscultatory blood pressure
measurement is preferred above using the oscillometric
method. In «Manual» Mode, the device serves only as a
pressure gauge. No oscillometric measurements will be
taken. The user can hear the systolic and diastolic Korotkoff
sounds by means of a stethoscope placed over the Brachial
artery.
20
Mode (cont.)
Start inflation – Press the Start/Stop Button to start inflation
of the cuff.
When maximum inflation pressure is reached, the ProBP
2400 will automatically begin linear deflation at a rate of
3mmHg/sec.
Re-inflate – Push and Hold the M Button during deflation
when the cuff pressure is below 200mmHg to re-inflate for as
long as the button is held (up to a max of 200mmHg) Release
the button to continue deflation. An additional push and hold
of the M Button will allow reinflation past 200mmHg to a max
of 299mmHg.