Measurement/Monitoring)........9 Mode -03- (Process Time Measurement).... 10 “How to Operate” Teaching Feature .....10 Congratulations on your purchase of a Shimpo DT-5TXR or DT-5TS digital panel mount tachometer. Whether Setting High and Low Limits ......... 11 measuring RPM or a complex function, we trust you will enjoy many years of professional results.
Never use wet hands when connecting or disconnecting a sensor or checking the unit. Items included with the DT-5TXR or DT-5TS are: • Mounting adapter, 2 pieces (attached) Always use the correct voltage (AC type: 85- • Mounting screw, 2 pieces (attached) 264VAC, DC type: 9-35VDC).
Display and Keypad s ’ s t i Secondary Display B Secondary display Negative Sign/Decimal Point Button Negative sign setting/decimal point shift Secondary Display A Secondary display Comparator Indicator Indicates comparator condition...
LED display will light up. diagram 1 Sensor Connection Use the chart below to determine how to connect the DT-5TXR/DT-5TS to Shimpo sensors: SENSOR TERMINAL NUMBER FREQUENCY or RPM RANGE OPERATION TEMPERATURE FILTER TYPE –...
4. Hold the panel meter in a horizontal position, insert it into the cut-out and press against gasket (diagram 5). 5. Replace the two mounting adapters (diagram 6). 6. The DT-5TXR/DT-5TS is secured on the panel by using the two mounting screws (provided) (see diagram 7). diagram 4...
NOTE: If a mistake is made entering any number, simply press the BACKSPACE button to erase the last digit on the primary display. Mode -01- (RPM or Rate Measurement) The following parameters must be set in mode -01- so that the DT-5TXR/DT-5TS will display the correct readings: FACTORY SETTING PARAMETER...
Page 8
Mode -01- (RPM or Rate Measurement) continued Set the parameters in mode -01- as follows: 7. Press the PAR button; the display will reflect: 1. Press the PAR button; the display will indicate: Parameter 4 is now showing on the primary display (1.0 second [factory setting]).
Mode -02- (Elapsed Time Measurement/Monitoring) The following parameters must be set in mode -02- so that the DT-5TXR/DT-5TS will display the correct readings: FACTORY SETTING PARAMETER DESCRIPTION RANGE DISPLAY MEANING Pulses p i t u – _ _ 0001 1 p/r...
Mode -03- (Process Time Measurement) The following parameters must be set in mode -03- so that the DT-5TXR/DT-5TS will display the correct readings: FACTORY SETTING PARAMETER SETTING RANGE DISPLAY MEANING Hours, minutes, seconds 00:00:00 0:00 Hundreds of a second 0:00...
NOTE: The DT-5TXR provides relay or transistor closures (see section “Modules Available for DT-5TXR” on page 15). Setting Functions The DT-5TXR/DT-5TS has a variety of functions that can be programmed to further customize the meter. Listed below are the available functions:...
Function 4 is now showing (factory setting). The function of the MEM button is to aid the NOTE: This function is mainly used with mode -01- DT-5TXR/DT-5TS to memorize and display through the only. secondary displays the maximum and minimum value of 8.
MODE -01- (RPM or Rate Measurement) Parameter 4 is selectable and is used when there is a The DT-5TXR/5TS units are fully scalable, which means that steady increase/decrease in speed and the operator wants RPM can be used to display linear speed in many units, to observe the changes (in this case higher update times i.e.
Page 14
DT-5TXR/DT-5TS telling it when the process started and also when it is finished upon the return of the press table. As was mentioned above, MODE -03- of the DT-5TXR/5TS is able to count and display that process time.
DT-5TXR. DT-5TXR-CPTR The model DT-5TXR-CPTR provides three “C” type relays that Most of the modules have either a “T” or “C” suffix in can be used for setting two “set points”, one high and the part number.
DT-5TXR with DOP-TRTR Module programming, but by making the proper connections. DT-5TXR-TRTR The DT-5TXR panel meter with the DOP-TRTR The connections for the DOP-FVTR module are as follows: module provides two high set points (limits) and two low set points in addition to the normal GO signal output.
Page 17
DT-5TXR-FVTR/DT-5TXR-FVC (continued) Instructions to set the analog output characteristics are as follows: 1. Press the FUN button eleven times; the display will reflect: 5. Press the FUN button again; the display indicates: The display shows that the maximum RPM is set for The display reflects the factory setting.
LO, the BCD output data goes to a high impedance state can be configured with a negative logic output (0=HI to facilitate the selection of one unique DT-5TXR-BCD unit and 1=LO) or a positive logic output (0=LO and 1=HI), among several that might be feeding the same PLC. At depending upon what the PC or PLC requires.
DImensions and Speci cations DT-5TXR MODEL Mode e t a r e t Elapsed Time Process Time Display Range 0:00:00 - 9:59:59 0:00:00 - 0:59:59 0 - 999999 with selectable decimal point 0.00 - 999.99 s ec 0.00 - 999.99 s ec...
This warranty shall not be effective if the product has been subject to overload, misuse, negligence, or accident, or if the product has been repaired or altered outside of Shimpo Instruments’s authorized control in any respect which in Shimpo Instruments’s judgment, adversely affects its condition or operation.