Set Up A Calculation - Hach Polymetron 9500 User Manual

Hide thumbs Also See for Polymetron 9500:
Table of Contents

Advertisement

5. Select an output mode option and push enter.
Option
ACTIVE
HOLD
TRANSFER
6. Select the sensors that will have their outputs (analog and relay) affected when one of the
discrete inputs becomes active. Push enter.
7. Use the arrows to select the ON DELAY value (the duration time delay between the discrete input
activation and the configured response of the controller). Push enter.
8. Use the arrows to select the OFF DELAY value (the duration time delay between the discrete
input de-activation and the configured response of the controller). Push enter.
9. Repeat steps 4–8 for each desired channel.
10. If a discrete input needs to be changed after initially set up:
a. Repeat steps 1–2 and an INPUT SETTINGS menu appears with the following options:
• CONTROL LOGIC
• SET WARNING
• OUTPUT MODE
• ON DELAY
• OFF DELAY
b. Select the desired option and push enter.
c. Make the desired changes and push enter to save the changes and return to the INPUT
SETTINGS menu.

Set up a calculation

Two sensors must be installed in the controller.
1. Push the menu key and select Polymetron 9500 SETUP>CALCULATION.
2. Select a menu option and choose from the displayed list or update the entry.
Option
SET VARIABLE X
SET PARAMETER X
SET VARIABLE Y
SET PARAMETER Y
SET FORMULA
DISPLAY FORMAT
SET UNITS
SET PARAMETER
3. Push enter to save the selection or setting and return to the CALCULATION menu.
Description
Output level continues to represent operating conditions.
Output level is held static.
Output level moves to a pre-configured value.
Description
Selects the sensor that corresponds to the variable set as "X".
Selects the measurement type that corresponds to the variable set as "X".
Selects the sensor that corresponds to the variable set as "Y".
Selects the measurement type that corresponds to the variable set as "Y".
NONE, X-Y, X/Y, [X/Y]%, [X+Y]/2, X*Y, [X-Y]%/X.
Sets the number of decimal places displayed in a calculation result.
Sets the units of measurement.
Assigns the name of the measurement.
English 37

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents