Introduction Thank you for purchasing the TD-01 Portable Indicator. Disclaimers Please read this document in its entirety before using the product to get the best performance and ensure safe and Information is given about products in this manual only proper operation.
Safety information This document describes the safety instructions for the operation of the digital indicator. Before operating the product, read this doc- ument carefully to familiarize yourself with the unit. Follow the instructions below to avoid risk of serious personal injury and V WARNING death.
Page 4
Safety information V CAUTION Follow the instructions below to avoid risk of personal injury or property damage. Put the unit power into standby before doing any of the following. o Installing load cells o Replacing batteries o Connecting other external equipment Never touch the connectors while the product is turned on.
Improper disposal of waste electrical/electronic Changes or modifications to this equipment not expressly equipment and batteries/accumulators can have seri- approved by TEAC CORPORATION for compliance could ous effects on the environment and human health void the user’s authority to operate this equipment.
Installing batteries Remove the back cover and insert the batteries. Button cell battery Close the cover after inserting the batteries The button cell battery is used to back up the clock. Misuse of batteries could cause them to rupture or leak, which might result in fire, Replace the button cell battery if the clock display shows that it is January 1, 2000 after the power is turned on, except for the injury or the staining of nearby materials.
1. Names and functions of parts 1-1. Front panel When Cal. Value Lock is OFF, press and hold to use the REC/REC MENU button Zero Balancing function (page 37). When Cal. Value Lock is ON, press this to forcibly set the indicator value to zero (Digital zero function).
1. Names and functions of parts LOAD: A/D converter plus over 1-2. Top panel −FULL: display minus over (less than the minimum display value) FULL: display plus over (greater than the maximum dis- play value) −OVER FULL: Input is outside the input range (neg- ative value) OVER FULL: Input is outside the input range (posi-...
1. Names and functions of parts 1-3. Sensor signal input terminals 1-4. Bottom panel Connect one sensor to either INPUT 1 or INPUT 2. Do not con- nect sensors to both inputs. TEDS(F) GND(G) +EXC(A) SIG(B) EXC(C) o Connector covers are attached to the connectors. Remove the connector covers to use the connectors.
1. Names and functions of parts 1-5. Screen transition diagram Press and hold Power supply ON o The last open screen will reappear. Press and hold Indicator value Graph Static Strain Disp. Mode Simple indication Memory and list selection/display Function Menu Press and hold Equivalent Input...
1. Names and functions of parts Power status 1-6. Home Screen : Operating on bus power from the USB port. The indicator value or graph display is shown after the unit is turned on. (The last open screen will reappear.) : Operating on batteries Use the g and t buttons to change screen views.
1. Names and functions of parts Press and hold 5+b Unit This is not indicated on the front panel, so it can be used to prevent unlocking when not desired. This is the set position (page 30). Press and hold 5+b to unlock this control lock. This can also be used to unlock when control lock High limit value has been set by pressing and holding b.
1. Names and functions of parts Waveform 1-7. Static Strain Disp. Mode Values between the low limit and the high limit are Set whether to show the input signal with the strain amount shown in green. unit (μst). Values above the high limit or below the low limit are Use when checking sensor output and unsteadiness in indica- shown in red.
2. Shortcut menus 2-1-2. Setting Memory 2-1. Memory and list selection/display Select the setting value memory. Press the READ button to open the following screen. This is the same setting screen as System settings 1 w Setting memory. See “10-1 Setting Memory” on page 55 2-1-1.
2. Shortcut menus 2-1-3. Indicator value list Mode Indicator recording mode 1: When REC key pressed (KEY) Indicator value list 2: When stability detected (AUTO) 3: When hold stops (ZONE) Sensor Sensor value memory number Indicator value Setting value list Indicator value h:m:s Time of recording...
2. Shortcut menus 2-1-4. Graph list Deleting data Graph list Graph analysis When the graph list is open, press and hold the REC and ESC buttons at the same time to delete the most recently recorded data. For other ways to delete data, see “11-3 Recorded data dele- tion”...
2. Shortcut menus Layered display of graphs When stability detected (AUTO) Indicator values will be recorded when stability is detected. Press the ENT button when the graph analysis screen is open. For stability detection methods, see “6-2 Motion Detect” on page 39. When hold stops (ZONE) Indicator values will be recorded when hold goes from ON to OFF.
2.5V, the sensor could be damaged. Button Connecting a TEDS sensor or one with 4 wires Strain gauge transducer TD-01 Portable Remove 7–8 mm of the covering from the wire being connected, and twist it so that the tip does (TEDS) TEDS Data not come apart.
3. Making connections 3-3. Interrupted wire checking Press and hold the ESC and ENT buttons to check for inter- rupted wires and show the results on the display. If the possibility of an interruption is detected, the location of the possible interruption will be shown in red. Interruptions can occur not only in strain gauges, but also in load cell cables.
4. Settings The following screen appears when you select 4-1. Basic operation Calibration w Equivalent Input Calibration. Press the MENU button to open the Function Menu. Selected item Changing a series of setting values. Depending on the menu item, setting screens might appear in a series.
4. Settings 4-2. Selecting setting values from 4-3. Inputting numerical setting values options The selected digit appears dark with a white background. The current value has a white background. Other options have gray backgrounds. Use the g and t buttons to change the selection, and use the 5 and b buttons to change the value.
4. Settings 4-4. Changing the decimal point 4-5. Lock position You can prohibit the changing of calibration and setting values. The decimal point position can only be changed for the Rated When locked, “Locked” appears with a white background at the Capacity.
4. Settings 4-6. Returning to the Home Screen 4-9. Setting menu list 4-9-1. Calibration Calibration Equivalent Input Calibration (page 32) Sensor value memory Rated Output Rated Capacity Zero Balancing Select EU Cal. Value Lock Actual Load Calibration (page 33) Sensor value memory Zero Balancing Use the g and t buttons to select TOP, and press the ENT Rated Capacity...
4. Settings 4-10. Setting value list 4-10-1. Calibration Cal. Setting Setting Unit Default Value value value Item Setting Format Setting range/options display value Lock lock memory Sensor value memory Options 1 to 6 Rated Output Input mV/V 3.000 0.300 to 5.000 ...
4. Settings 4-10-5. System settings 1 Cal. Setting Setting Unit Default Value value value Item Setting Format Setting range/options display value Lock lock memory Setting value memory Options 1 to 6 D/A Converter D/A output Options ON, OFF D/A Zero Input Set unit 000.00...
5. Calibration Sensor check before calibrating Connecting the unit with a strain gauge transducer and setting how the indicator values will be shown is called "cali- After connecting a sensor and turning the unit on, calibration bration". The following three calibration methods can be used is not possible if the indicator value is unstable or an error with the unit.
5. Calibration Use the 5b buttons to change the selection, and 5-1. Procedures shared by all cali- select in order System Settings 1 w Lock bration methods The three calibration methods are equivalent input calibra- tion, actual load calibration and TEDS calibration. All the calibration methods have the same procedures before and after calibration.
5. Calibration Press the ENT button to confirm the selection. 5-2. Equivalent input calibration This method determines the calibration value by recording rated output and rated capacity values from a test results table. Use this to calibrate easily when an actual load cannot be applied.
5. Calibration 5-2-3. Rated Output 5-3. Actual Load Calibration Input range: 0.300–5.000 mV/V Set the rated output of the strain gauge transducer being Calibrate by applying an actual load to the sensor. used. Sensor value memory 5-2-4. Rated Capacity selection Set the rated capacity of the strain gauge transducer being used.
5. Calibration 5-3-3. Zero Balancing 5-3-5. Select EU With no load on the sensor, press the ENT button. Select the unit that corresponds to the indicator value. During zero balancing, a pop-up appears to show that it is in progress. When zero balancing completes, the setting screen for the next procedure appears.
5. Calibration 5-4-2. Rated Output/Rated Capacity display 5-4. TEDS calibration After reading the TEDS data, the Rated Output/Rated Capacity is shown. A TEDS sensor has calibration information, including rated out- put and rated capacity, stored in its memory. Rated capacity digit shown during TEDS calibration TEDS calibration reads this calibration information to automati- cally record calibration values.
5. Calibration 5-4-4. Select EU 5-6. Select Min. Grid Options: N, kN, kPa, MPa, g, kg, sht, ton, mN·m, N·m, kN·m, none Select the unit that corresponds to the indicator value. Set the minimum digital change of the indicator value. o If a sensor with TEDS built in is connected, the read unit will be set.
5. Calibration 5-8. Select EU 5-10. Zero Balancing Select the indicator value unit. You can conduct Zero Balancing calibration even without Select the unit that corresponds to the indicator value. resetting the calibration value. Set Cal. Value Lock to OFF. See “5-1-1 Locking and unlocking calibration values”...
6. Condition Settings 6-1. Filter 6-1-1. Select Num. of Moving Avg. Set the moving average number for the measured data. Options: OFF, 16, 32, 64, 128, 256, 512, 1024, 2048...
6. Condition Setting 6-2. Motion Detect Set the parameters used to detect stability. The indicator value is judged to be stable if a difference less than the set width occurs between the current indicator value and the indicator value 100 mSec before, and this continues for the set time. The Motion Detect function is closely related to the comparison mode.
6. Condition Setting 6-3. Zero Tracking This function automatically tracks drift and other gradual changes to the zero point. With Zero Tracking, when the zero point moves no more than the set width, the indicator value is automatically adjusted to zero each set time interval.
6. Condition Setting 6-5-2. Digital Zero Limit Value 6-4. Digital Zero Offset Set the range for Digital Zero capture. (The setting value unit is the same as for the indicator value.) The set value is subtracted from the measured value (digital zero).
6. Condition Setting 6-6. Select Data Output Select the data output from the D/A Converter. Select Data Output has the following two options. The hold value is output in coordination with the display. The input is output as is.
7. Comparison Setting 7-1. Comp. Value Setting 7-2. Comp. Mode Setting Use this function to set the high limit and low limit values, Set the conditions for judging high limit and low limit. compare them with indicator values, and turn judgment dis- play ON for each one.
7. Comparison Setting 7-3. Hysteresis This function applies a width to switching from ON to OFF for high limit and low limit judgment display. If the indicator value fluctuates near a value, causing judgment display to chatter between ON and OFF, adjustment of the hysteresis value can prevent this chattering.
7. Comparison Setting 7-4. Nearly Zero Set the range in which indicator values are evaluated as being nearly zero. NOTE The Nearly Zero judgment result is closely related to high limit and low limit judgment display. See “7-2 Comp. Mode Setting” on page 43 for details.
8. Hold Function Setting 8-1. Hold Mode Hold Mode has four settings. No hold SAMPLE Sample and hold PEAK Peak hold BOTTOM Bottom hold NOTE Press and hold the ENT button to turn hold on/off. The hold status is shown at the bottom of the display.
8. Hold Function Setting 8-1-1. Sample and hold Press and hold the ENT button to hold the indicator value. Press and hold the ENT button to hold the indicator value, and press and hold it again to stop holding the value. Indicator value Sensor input value...
8. Hold Function Setting 8-1-2. Peak hold 8-1-2-1. No zone definition After pressing and holding the ENT button to turn hold on, the maximum value (peak value) in the indicator value positive direction is shown held. Press and hold the ENT button again to turn hold off, clearing the peak hold. When holding, the peak value is also reset by pressing the CLEAR (ESC) button.
8. Hold Function Setting 8-1-2-2. Zone definition used The indicator value when the hold operation completes continues to be shown. When holding, the peak value is reset by pressing the CLEAR (ESC) button. Indicator value Indicator value Sensor input value When zone definition Time When holding...
8. Hold Function Setting 8-1-3. Bottom hold 8-1-3-1. No zone definition Press and hold the ENT button to turn hold on and hold and display the minimum (bottom) indicator value. Press and hold the ENT button again to turn hold off, clearing the bottom hold. When holding, the bottom value is also reset by pressing the CLEAR (ESC) button.
8. Hold Function Setting 8-1-3-2. Zone definition used The indicator value when the hold operation completes continues to be shown. When holding, the bottom value is reset by pressing the CLEAR (ESC) button. Indicator value Indicator value Sensor input value When zone definition Time When holding...
8. Hold Function Setting 8-2. Zone Definition When set to ON, the indicator value will continue to be shown after the hold ends. Pressing the CLEAR (ESC) button will clear the held value. Options: ON, OFF 8-3. Auto Zero Set whether or not to automatically execute a Digital Zero when a hold starts (ON/OFF).
9. Graph settings 9-1. Time axis setting (X) 9-3. Trigger mode Set the width of the time axis. Enable/disable trigger operation Trigger disabled 9-2. Indicator value axis setting (Y) Trigger enabled Set the range of indicator values displayed. 9-4. Trigger 9-4-1.
9. Graph settings 9-4-2. Trigger level Set the indicator value level that causes trigger operation. The trigger point is when the indicator value reaches the set trigger level value from the direction set by the trigger polarity. 9-4-3. Trigger position This sets the display position of the cursor.
10. System settings 1 10-1. Setting Memory 10-2. D/A Converter You can save up to 6 memories with setting values and switch The D/A Converter allows for analog output that corresponds among them. to the unit indicator value. The analog output range can be set to either ±1V or ±2V volt- age output.
10. System settings 1 10-2-2. D/A Zero Setting example 2 Set the indicator value that you want to output for D/A Zero D/A Zero 020.00 (0V voltage). D/A Full Scale 100.00 Indicator value D/A output 120.00 20.00 −80.00 −1 V Setting example 3 10-2-3.
10. System settings 1 10-2-4. D/A Max. Voltage 10-3. Lock Set the maximum D/A voltage. The load also uses the same output voltage range. For example, when set to 1V, +1 V will be 10-3-1. Cal. Value Lock output when the “10-2-2. D/A Zero” and “10-2-3. D/A Full Scale” setting values are added.
10. System settings 1 10-4. Brightness 10-5. Auto power off Adjust the brightness of the LCD screen backlight and the time Set the amount of time without any button operations until until it dims. the unit enters standby. 10-4-1. Brightness Select OFF, 2 min., 5 min.,10 min, 30 min or 60 min.
10-9-2. Date & time setting 10-8. Device number Set the current time for the clock. You can give this unit a unique number. Do this when using multiple TD-01 Portable units, for example. o Press the ENT button after setting all items.
11. System settings 2 11-1. Battery type 11-3. Recorded data deletion Set the type of AA batteries. This is used to estimate remaining Saved indicator values and graph data can be deleted. battery charge. You can select which data is deleted first using “Indicator” and “Graph”...
12. TEDS Settings By connecting a sensor that supports IEEE1451.4 Transducer 12-1. TEDS Data Electronic Data Sheets (TEDS) to the unit, the Rated Output stored in the sensor can be loaded and this can be applied to Select TEDS Data to show the following items. calibration of the indicator with this function.
13. Error message list Message Explanation LOAD ADC plus over −LOAD ADC minus over FULL Display plus over (greater than maximum display value) −FULL Display minus over (less than minimum display value) OVER FULL Input is exceeding maximum input range (5.0 mV/V) −OVER FULL Input is less than minimum input range (−5.0 mV/V) ZERO OVER...
Need help?
Do you have a question about the TD-01 Portable and is the answer not in the manual?
Questions and answers