Specifications; Electrical Specifications; Table 2: Operating Mode - Tektronix AFG3000 Series Technical Reference

Arbitrary function generators specifications and performance verification
Hide thumbs Also See for AFG3000 Series:
Table of Contents

Advertisement

Specifications
Electrical specifications

Table 2: Operating mode

Characteristic
Run mode
Burst count
Internal trigger rate
AFG3000 Series Specifications and Performance Verification
These specifications apply to all AFG3000 Series Arbitrary Function Generators.
All specifications are guaranteed unless labeled "typical". Typical specifications
are provided for your convenience but are not guaranteed.
Specifications that are check marked with the
indirectly) in the Performance Verification section.
All specifications apply to the arbitrary function generator unless noted otherwise.
These specifications are valid under the following conditions:
The instrument must have been calibrated/adjusted at an ambient temperature
between +20 °C and +30 °C.
The instrument must be operating at an ambient temperature between 0 °C
and +50 °C.
The instrument must have had a warm-up period of at least 20 minutes.
The instrument must be in an environment with temperature, altitude, and
humidity within the operating limits described in these specifications.
Description
Continuous, Modulation, Sweep, and Burst
1 to 1,000,000 cycles or infinite
1.000 μs to 500.0 s
symbol are checked directly (or