Configure The Setup Table - DENT Instruments ELITEpro XC Operator's Manual

Portable energy data logger with elog 15 software
Hide thumbs Also See for ELITEpro XC:
Table of Contents

Advertisement

C
ONFIGURE THE
The setup table consists of several sections where the attributes of the ELITEpro XC are defined.
Logger Type
ELITEpro XC is the default
entry. This field cannot
be edited once you are in
the Setup Table window.
Logger type is chosen
from the Select A Setup
Table Type window.
1.800.561.8187
S
T
ETUP
ABLE
Setup Table Name
Enter a name for the
setup table. 20 character
limit. This is helpful for
identifying different
setup tables. NOTE: The
setup table name need
not be the same as the
.sut file name.
www.
Data Interval
The default integration
Auto, 60 Hz, or 50Hz.
period is fifteen minutes.
Default is 60 Hz.
The choices are 1, 3, 15,
30 seconds; 1, 2, 5, 10,
15, 20 and 30 minutes; 1
and 12 hours; 1 day. This
tells the logger at what
time intervals data is to
be stored in the memory.
For example, if the
integration period is set
for 30 minutes and
average watts is being
monitored, then every 30
minutes the logger
records the average
power use (watts) for
that channel over the
preceding 30 minute
interval, based on
approximately 14,400
measurements of the
monitored power draw. If
Maximum (and/or
Minimum) values are
being recorded the
highest (and/or lowest)
of those 14,400 readings
are also saved.