Pressure Equipment Directive, PED, 97/23/EC

Siemens Level Transmitters with flanged, threaded, or sanitary clamp type process
mounts have no pressure-bearing housing of their own and, therefore, do not come under
the Pressure Equipment Directive as pressure or safety accessories, (see EU Commission
Guidelines 1/8 and 1/20).

Mounting location

Notes:
• Correct location is key to a successful application.
• Avoid reflective interference from vessel walls and obstructions by following the
guidelines below

Nozzle design

Threaded PVDF Antenna
Min. dia. 50 mm (2")
The end of the antenna must protrude a minimum of 10 mm (0.4") to avoid false
echoes being reflected from the nozzle.
Minimum recommended nozzle diameter for the threaded PVDF antenna is 50 mm
(2").
An antenna extension (100 mm/ 3.93") is available for any version except the
threaded PVDF antenna.

Nozzle location

Avoid central locations on tall, narrow vessels.
Environment
Provide an environment
suitable to the housing rating
and materials of construction.
Provide a sunshield if the
instrument will be mounted in
direct sunlight.
Page 20 SITRANS LR250 (PROFIBUS PA) – OPERATING INSTRUCTIONS
Stainless Steel Horn Antenna
Min.
clearance:
10 mm (0.4")
Min.
clearance:
10 mm (0.4")
preferred
undesirable
ambient temperature
–40 °C to +80 °C
(–40 °F to +176 °F)
process temperature
- with FKM O-ring:
–40 to +200 °C
(–40 to +392 °F)
- with FFKM O-ring:
–20 to +200 °C
(–4 to +392 °F)
7ML19985JF04