9 Technical data

9.2. More technical data

9.2.1. testo 770-1/-2

Feature
DC voltage
2
3
4
AC voltage
,
,
DC current
- jaws [A]
- jack [µA]
(testo 770-2)
3
AC current
5
- jaws [A]
- jack [µA]
2
4
(testo 770-2)
,
Resistance
Continuity alarm
Diode test
Capacity
1
The lower measuring ranges are only specified from 5% (does not apply
to DC current / AC current measurements with the current probe)
2
Signal bandwidth 40 Hz to 1 kHz
3
In the case of a mixed signal (AC + DC), only the purely AC component is
taken into account
4
As the frequency increases (over 400 Hz), the accuracy deteriorates
+/- (1.5% of m.v. + 3 digits) for 400Hz to 750Hz / +/- (2.0% of m.v. + 3digits)
for 750Hz to 1kHz
5
Frequency of AC current up to 400 Hz
6
Specification is valid for capacitances > 10 nF
20
Measuring
Resolution Accuracy
1
range
4.000 V
40.00 V
400.0 V
600 V
4.000 V
1 mV
40.00 V
10 mV
400.0 V
100 mV
600 V
1 V
400 A
400 µA
400 A
400 µA
400.0 Ohm
0.1 Ohm
4.000 kOhm
1 Ohm
40.00 kOhm
10 Ohm
400.0 kOhm
100 Ohm
4.000 MOhm
1 kOhm
40.00 MOhm
10 kOhm
<0 to 30 Ohm
yes (0 to 2.5 V)
6
51.20 nF
0.01 nF
1 mV
± (0.8% of meas. val. +
10 mV
3 digits)
100 mV
1 V
± (1.0% of meas. val. +
3 digits)
0.1 A
± (2.0% of meas. val. +
5 digits)
0.1 µA
± (1.5% of meas. val. +
5 digits)
0.1 A
± (2.0% of meas. val. +
5 digits)
0.1 µA
± (1.5% of meas. val. +
5 digits)
± (1.5% of meas. val. +
3 digits)
± 10% typically

Hide quick links: