810
Users Manual
Display Message
Failed to read RPM from
Tachometer
Sensor not connected
Failed to read Sensor type
Cable is open
Cable is short
Data acquisition failed
Measurement storage failed
Measurement time out
B-2
Table B-1. Vibration Tester Warning and Error Messages (cont.)
Tachometer cannot read the RPM of the machine. Follow the correct procedure to measure
the RPM.
The Tester cannot find the Sensor. See "How to Connect the Sensor" to make the
connections.
The Tester cannot find the Sensor. See "How to Connect the Sensor" to make the
connections.
The Tester cannot find the Sensor. See "How to Connect the Sensor" to make the
connections.
Cable or Sensor is bad. Change the cable or Sensor. Contact Fluke for a new Sensor.
The Sensor mount method or measurement procedure is incorrect. See "How to Take a
Measurement" and "Sensor Mounting" for the correct procedure.
The Tester memory is full. Make a data backup with the Viewer software and remove data
from the Tester memory.
Allowed measurement time has been exceeded. The time to make a measurement is
30 minutes with triaxial Sensor and 45 minutes with single-axis Sensor. Make a new
measurement within the allowed time.