Page 3
Essential Instructions Read this page before proceeding! Emerson designs, manufactures and tests its products to meet many national and international stan- dards. Because these sensors are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount products.
Page 4
About This Document This manual contains instructions for installation and operation of the Rosemount 3200HP pH Sensors. The following list provides concerning all revisions of this document. Rev. Level Date Notes 04/17 Reformatted to reflect the latest Emerson documentation style...
Instruction Manual Specifications LIQ-MAN-3200HP April 2017 Section 1: Specifications Specifications Table 1-1: Rosemount 3200HP sensor specifications Conductivity Range > 0.4 S/cm Sample Temperature 0 °C to 70 °C (32 °F to 160 °F) Sample Inlet Pressure 5 to 10 psig (134-170 kPa abs); no back pressure allowed; sample must drain to open atmosphere Accuracy ±...
Specifications Instruction Manual April 2017 LIQ-MAN-3200HP Product Certifications Please see online certificates for further details. IECEx Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C) Per standards IEC60079-0 : 2011, IEC 60079-11 : 2011 ATEX II 1 G Ex ia IIC T4 Ga (-20˚C ≤ Ta ≤ +60 °C) Per standards EN 60079-0: 2012+A11:2013, EN 60079-11:2012 Intrinsically Safe for use in Class I, II, and III, Division 1, Groups A, B, C, D, E, F, and G;...
Connect the sample and drain lines to the flow cell. Process connections are ¼ inch FNPT. Note: The flow cell must drain immediately to open atmosphere. Do not use a long or convoluted drain line. Figure 2-1: Dimensional Drawing for Rosemount 3200HP Installation...
April 2017 LIQ-MAN-3200HP Figure 2-2: Rosemount 3200HP Putting the Rosemount 3200HP Sensor in Service Wire the cable to the transmitter. See the wiring diagrams in the wiring section. Start the sample flow to fill the flow cell with water. Unscrew the plastic nut on the flow cell and slip the nut over the sensor. See Figure 2-1.
Page 11
60° from vertical. Watch for bubbles leaving the port. Bleed an additional 2 mL of solution after the last bubble has escaped. Tighten the bleed screw. Figure 2-4: Rosemount 3200HP Figure 2-3: Rosemount 3200HP Figure 2-5: Rosemount 3200HP...
Page 12
April 2017 LIQ-MAN-3200HP Connect the VP cable to the sensor. The sensor Figure 2-7: Rosemount 3200HP connector is keyed to ensure proper mating with the cable receptacle. Once the key has slid into the mating slot, hand tighten the connection by turning the knurled ring clockwise.
LIQ-MAN-3200HP April 2017 Section 3: Wiring Wiring for Rosemount 3200HP The Rosemount 3200HP sensor is available with Variopol (VP) quick disconnect cable only. For other wiring diagrams not shown below, please refer to the Liquid Transmitter Wiring Diagrams. Table 3-1: VP6 and VP8 Cable Cable Length 10 ft.
Page 14
Wiring Instruction Manual April 2017 LIQ-MAN-3200HP Figure 3-2: Rosemount 3200HP Sensor Wiring to Rosemount 1057 Transmitter Installation...
Page 15
Instruction Manual Wiring LIQ-MAN-3200HP April 2017 Figure 3-3: Rosemount 3200HP Sensor Wiring to Rosemount 1066 Transmitter Figure 3-4: Rosemount 3200HP Sensor Wiring to Rosemount 5081 Transmitter Installation...
Page 16
Wiring Instruction Manual April 2017 LIQ-MAN-3200HP Installation...
Instruction Manual Calibration LIQ-MAN-3200HP April 2017 Section 4: Calibration Calibration Select two stable buffer solutions. Ideally, the calibration buffers should bracket the expected pH of the sample and be at least two pH units apart. Use the ring at the bottom of the back plate to hold the beaker containing the buffer.
Page 18
Calibration Instruction Manual April 2017 LIQ-MAN-3200HP Calibration...
Instruction Manual Maintenance LIQ-MAN-3200HP April 2017 Section 5: Maintenance Replacing the Electrolyte Solution (PN 9210391) The reference electrolyte solution should last two to three months. To replace the electrolyte solution: Remove the reservoir bottle from the clip on the back plate. Turn the bottle upright. Unscrew the cap without disconnecting the filter or electrolyte tube.
Page 20
Maintenance Instruction Manual April 2017 LIQ-MAN-3200HP Fill the reference junction port with electrolyte solution or deionized water until the level slightly overflows the port. Place the new capillary disc with the sealing O-ring facing down in the port. Use your finger tips –...
1. Remove air bubbles. Refer to step 11 in the Putting the error blocking flow or the inside membrane on Rosemount 3200HP Sensor in Service (Section 2.3). the capillary disc is fouled with biological 2. Discard the fill solution in the reservoir and sensor. Rinse the growth.
Page 22
Troubleshooting Instruction Manual April 2017 LIQ-MAN-3200HP Troubleshooting...
Instruction Manual Accessories LIQ-MAN-3200HP April 2017 Section 7: Accessories Accessories Table 7-1: Accessories for Rosemount 3200HP sensor Part Number Description 24281-00 15 ft. cable with mating VP8 connector 24281-01 25 ft. cable with mating VP8 connector 24281-02 2.5 ft. cable with mating VP8 connector 24281-03 50 ft.
Page 24
Accessories Instruction Manual April 2017 LIQ-MAN-3200HP Accessories...
Instruction Manual EC Declaration of Conformity LIQ-MAN-3200HP April 2017 EC Declaration of Conformity Note: Please see website for most recent Declaration. EC Declaration of Conformity...
Page 26
EC Declaration of Conformity Instruction Manual April 2017 LIQ-MAN-3200HP EC Declaration of Conformity...