Figure 5-1: Replacing the capillary disc
14
Fill the reference junction port with electrolyte solution or deionized water until the level
slightly overflows the port.
Place the new capillary disc with the sealing O-ring facing down in the port. Use your finger
tips – do not let your bare fingers touch the white filter membrane – to push the capillary
disc straight into the port. See
end of a clean ¼ inch compression fitting nut on the transparent portion of the disc and
push the disc into place by gently press straight down on the nut.
Reconnect the VP cable to the sensor. Place the sensor back in the flow cell and reattach
the electrolyte tube to the reservoir filter.
Slide the reservoir back into the clip.
Bleed air bubbles from the tube and sensor. Refer to step 11 in the
3200HP Sensor in Service (Section 2.3)
Figure
5-1. If inserting the disc is difficult, place the open