Clean the electrode in a mild detergent solution. Wipe the
probe with a soft tissue paper. Avoid touching the glass
membrane with your fingers. Wash thoroughly in tap water
and then in distilled water. Recalibrate the meter after
cleaning the electrode.
6

TROUBLESHOOTING

Problem
Power ON but no Display
'LO' displays in the LCD
Unstable Reading
'Er1' Display
Not able to Calibrate
15 of 21
Cause
a) Batteries not in place.
a) Low battery.
a) Electrode not deep
enough in sample.
b) Dirty electrode.
c) Broken electrode.
a) Buffer value out of
tolerance.
a) Display freezes.
b) Faulty electrode.
Solution
a) Insert batteries.
b) Re-insert batteries in
correct polarity.
a) Replace batteries with
fresh ones.
a) Place electrode deeper in
sample.
b) Clean electrode and recali-
brate.
c) Replace electrode.
a) Use new buffer solution
and recalibrate.
a) Release reading by
pressing HOLD key.
b) Replace electrode.
INST.21