Recommended Separation Distances Between Portable and Mobile RF Communications Equi-
The AMM213TD/AMM240WTD system is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The user of the AMM213TD/AMM240WTD system can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the AMM213TD/AMM240WTD system as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power (W) of transmitter
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in me-
ters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
pement and the AMM213TD/AMM240WTD System
Separation distance (m) according to frequency of transmitter
150kHz to 80MHz
d=1.17, p
0.12
0.37
1.17
3.70
11.70
80MHz to 800MHz
d=1.17, p
0.12
0.37
1.17
3.70
11.70
800MHz to 2.5GHz
d=2.33, p
0.23
0.74
2.33
7.37
23.30