9 Supplement

9.1 Technical data

General data
Material 316L corresponds to 1.4404 or 1.4435
Materials, wetted parts
Process fitting - thread
-
Process fitting - flange
-
Process seal
-
Instrument seal
-
Antenna cover
-
Cover - Mounting adapter (optional)
-
Seal - Mounting adapter (optional)
-
Materials, non-wetted parts
Housing plastic
-
Aluminium die-casting housing
-
Stainless steel housing - precision cast-
-
ing
Stainless steel housing, electropolished
-
Seal between housing and housing
-
cover
Ground terminal
-
Mounting adapter (option)
-
Sensor length
Instrument weight (depending on process
fitting)
Process fittings
Pipe thread, cylindrical (ISO 228 T1)
-
American pipe thread, tapered
-
Flanges
-
Frequency range
Measuring range
Angle of aperture of the antenna 3 dB
Encapsulated horn antenna (G1½ A)
-
Plastic encapsulated antenna with PP
-
cover
Output variable
Output
VEGAMIP R61 • - Relay
316L
316L
Klingersil C-4400
FKM (Viton)
PP, PTFE
0
Ceramic Al
2
3
Graphite
plastic PBT (Polyester)
Aluminium die-casting AlSi10Mg, powder-coated -
basis: Polyester
316L
316L
NBR (stainless steel housing, investment casting),
silicone (Aluminium/plastic housing, stainless steel
housing, electro-polished)
316L
316L
See chapter "Dimensions"
0.8 ... 4 kg (0.18 ... 8.82 lbs)
G1½ A
1½ NPT
DIN from DN 50, ANSI from 2"
K band, 24.085 GHz (ISM band)
0.1 ... 100 m (0.33 ... 328 ft)
20 °
10 °
Relay output, 2 floating spdts