Managing Protocols In The Instrument - Maico Ero Scan Pro Operating Instructions Manual

Table of Contents

Advertisement

Operating Instructions

9.2 Managing Protocols in the Instrument

Remove Protocol or Series
To remove protocols or series from the instrument, select the protocols or series to be removed and click
Remove
the
button located in the right panel of the Protocols in Instrument window. Use ctrl+click or
shift+click to select multiple protocols to remove or load.
Manufacturer default protocols cannot be removed – use the hide button instead.
When series are removed from the instrument they are DELETED permanently.
To load new or modified protocols into the
instrument or to remove unused protocols from
the instrument, click on the button located in
the upper right corner of the main window
Manage Protocols and Protocol Series
This dialog allows the user to manage the
protocols and series in the instrument. On the
right panel of the window are the following
buttons:
View: displays the selected protocol or series
settings
Remove: removes the selected protocol or
series from the instrument
Create New Series: opens the "Create New
Series" dialog shown in figure 85.
Instructions for creating a new series of
protocols can be found in section 9.3 of
this manual.
Load: opens the directory from which the user
can select a protocol file to load in the
instrument
Done: exit
47
.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents