  • Page 1 SP-360 Handheld Fluorometer - Fluorescein Operation Manual Rev. A Firmware version 1.xx...
  • Page 4: General Description

    1 General Description 1.1 Specification Fluorescein Measurement Range 0 to 600 ppb Excitation Wavelength 470 nm LED Emission Wavelength 520 nm Wavelength Accuracy ±1 nm Resolution 0.1 ppb Accuracy ±1% for > 50 ppb, 0.1 ppb for < 10 ppb Calibration Solution Point 0, 50, 250 and 500 ppb Others Battery...
  • Page 5 • • • • Figure 1 Light Shield in the Open Position...
  • Page 7 MEASUREMENT Fluorescein 360.0 Calibration Info Key Functions > < Keys...
  • Page 8 MEASUREMENT FLUORESCEIN CALIBRATION Insert DI Solution Press Zero Button Zero Slope Exit...
  • Page 9 MEASUREMENT FLUORESCEIN CALIBRATION Insert DI Solution √ Click Zero Button Insert ppb Solution Click Slope Button Press cycle to choose solution Cycle Slope Exit MEASUREMENT FLUORESCEIN CALIBRATION Insert DI Solution √ Click Zero Button Insert ppb Solution √ Click Slope Button Calibration Succeed Press Continue to Proceed Cycle...
  • Page 10 MEASUREMENT FLUORESCEIN CALIBRATION Insert 50 Solution √ Click Slope Button Insert ppb Solution Click Slope Button Cycle Slope Exit...
  • Page 11 MEASUREMENT FLUORESCEIN CALIBRATION Insert 50 Solution √ Click Slope Button Insert ppb Solution √ Click Slope Button Calibration Succeed Press Continue to Proceed Cycle Slope Exit MEASUREMENT FLUORESCEIN CALIBRATION Insert 250 Solution √ Click Slope Button Insert ppb Solution Click Slope Button Cycle Slope Exit...
  • Page 12 MEASUREMENT FLUORESCEIN CALIBRATION Insert 250 √ Solution √ Click Slope Button Insert ppb Solution √ Click Slope Button Calibration Succeed Slope Cycle Exit...
  • Page 14 MEASUREMENT MEASUREMENT DEVICE INFORMATION DEVICE INFORMATION [1] 4 Serial Number 170001 [2] 470 [5] 130 Hardware Ver [3] 340 Software Ver 1.1.1 Software Rev [4] 190 Battery Satus WiFi SSID SP360-170001 FLRC Calib Factory Reset Diagnosis Exit Exit...