Preparing Field Current Cable C, Connection To Measuring Sensor - Honeywell TWM 9000 Quick Start Manual

Electromagnetic flow converter
Table of Contents

Advertisement

TWM 9000

3.4.10 Preparing field current cable C, connection to measuring sensor

INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
• Field current cable C is not part of the scope of delivery.
• The shield for field current cable C can be connected to the measuring sensor.
• Bending radius: ≥ 50 mm / 2"
Required materials
• Heat-shrinkable tubing
• 3 wire end ferrules to DIN 46 228: size according to the cable being used
Figure 3-12: Field current cable C, preparation for the measuring sensor
a = 50 mm / 2"
1 Strip the conductor to dimension a.
2 Remove any shield that is present.
3 Pull a shrinkable tube over the prepared cable.
4 Crimp the wire end ferrules onto the conductors 7, 8 and 9.
34-VF-25-53 iss.3 GLO Aug 10 US