TWM 9000
Wall-mounted housing
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
• The connection of the outer shield (60) is carried out in the wall-mounted housing via the
stranded drain wire (6).
• Bending radius: ≥ 50 mm / 2"
Required materials:
• Push-on connector 6.3 mm / 0.25", insulation to DIN 46245 for conductor Ø = 0.5...1 mm
AWG 20...17
• PVC insulation tubing, Ø2.5 mm / 0.1"
• Heat-shrinkable tubing
• Wire end ferrule to DIN 46 228: E 1.5-8 for the stranded drain wire (1)
• 4 wire end ferrules to DIN 46 228: E 0.5-8 for insulated conductors 2 and 3 and the stranded
drain wires (20, 30)
Figure 3-7: Signal cable B, preparation for wall-mounted housing
a = 80 mm / 3.15"
1 Strip the conductor to dimension a.
2 Cut off the inner shield (10), the outer shield (60) and the shields for the conductor (2, 3). Make
sure not to damage the stranded drain wires (1, 6, 20, 30).
3 Slide the insulation tubing over the stranded drain wires.
4 Crimp the push-on connector onto the stranded drain wire (6).
5 Crimp the wire end ferrules onto the conductors and stranded drain wires (1, 20, 30).
6 Pull the heat-shrinkable tubing over the prepared signal cable.
34-VF-25-53 iss.3 GLO Aug 10 US
ELECTRICAL CONNECTIONS
www.honeywell.com/ps
3
2
/