4.1

Installation site requirements

• The instrument requires an installation area of approx. 250 x 700 mm. A clearance of at least
25 cm must be left between the instrument and the closest wall or other instruments.
• If the stainer is to be operated with rinsing stations, fresh water and waste water connections
must be no more than 1.2 m from the ports on the back of the instrument.
• It must be ensured that there is sufficient clear space (approx. 30 cm) above the instrument
to allow unobstructed opening of the optional hood.
• Room temperature consistently between +10 °C and +30 °C.
• Relative air humidity maximum 80 %, non-condensing.
• Practically vibration-free floor. Avoid vibrations, direct sunlight and heavy variations in
temperature.
• The chemicals used in the Leica ST4020 are easily inflammable and hazardous to
health.
• The installation site for the Leica ST4020 must be well ventilated, and there must be
no ignition sources of any kind in the area.
• The instrument must not be operated in areas at risk of explosion.
4.2

Unpacking the instrument

First check the shipment for external damages upon arrival. If it is evident that the
shipment was damaged during transport, please make a claim to the carrier imme-
diately.
• Open the packaging.
• Remove foam material.
• Take out all accessories and the instructions for use.
• Check the contents of the delivered pack against the order form to ensure that all components
and accessories are included.