Fluke 435-II Service Manual page 43

Three phase energy and power quality analyzer
Hide thumbs Also See for 435-II:
Table of Contents

Advertisement

Operate
F5
Press
to leave the SETUP DATE menu.
4. Select the calibration mode.
The Calibration Adjustment Procedure uses built-in calibration setups, that can be
accessed in the calibration mode.
To enter the calibration mode proceed as follows:
Press and hold
The display shows the
The display shows the calibration step
or
:IDLE (valid)
:IDLE (invalid)
Continue as indicated in section 5.2.
You can leave the calibration mode without changing the calibration data by turning the
Analyzer off.
Explanation of screen messages and key functions.
When the Analyzer is in the calibration Mode, only the
and the
The calibration adjustment screen shows the actual calibration step (name and number)
and its status:
Cal Name
Cal Name
(CL nnnn)
can be:
Status (...)
IDLE (valid)
IDLE (invalid)
BUSY aaa% bbb%
READY
to adjust each selected item Year, Month, and Day.
, press and release
screen (Calibration Adjustment mode).
MAINTENANCE
Warming Up (CL 0200)
, and the soft key menu.
key can be operated, unless otherwise stated.
(CL nnnn) :Status (...)
Name of the selected calibration step, e.g.
Number of the calibration step
After (re)entering this step, the calibration process is not started.
The calibration data of this step are valid. This means that the
last time this step was done, the calibration was successful. It
does not necessarily mean that the unit meets the specifications
related to this step!
After (re)entering this step, the calibration process is not started.
The calibration data are invalid. This means that the last time this
step was done, the calibration was not successful. Most probably
the unit will not meet the specifications if the actual calibration
data are saved.
Calibration adjustment step in progress; progress % for Input A
and Input B. During Warming Up the elapsed time is shown.
Calibration adjustment step finished.
5.3 Starting The Calibration
F5
, then release
, the calibration status
F5
F1
to
Warming Up
Calibration Adjustment
.
soft keys, the
5
5-5