Fluke 435-II Service Manual page 30

Three phase energy and power quality analyzer
Hide thumbs Also See for 435-II:
Table of Contents

Advertisement

Fluke 434-II/435-II/437-II
Service Manual
1. Set the Analyzer to 120 V, 60 Hz nominal (SETUP menu, arrow and ENTER keys).
2. To check the A/L1, B/L2, C/L3 inputs: connect the N input to Ground (See Figure 4-
3. Set the Calibrator to 60 Hz, 60 V and then to OPR.
4. Fluke 434-II: check for a voltage readout V rms between 59.4 ... 60.6 V.
5. Check the 120 V range according to the table below.
Set Calibrator to
60 V, 60 Hz, OPR
120 V, 60 Hz, OPR
240 V, 60 Hz, OPR
6.
Optional Test. Bandwidth Check of Channel A/L1, B/L2, C/L3 (*):
7. Set the Calibrator to 120 V, 60 Hz and OPR.
8. Now check the voltage channels one by one. Use the Calibrator's field edit function
9. Increase the frequency to 3 kHz.
Voltage Channel to be verified
A/L1
B/L2
C/L3
10. Set the Calibrator to STBY.
4-10
Press MENU, select Volts/Amps/Hertz, press F5 - OK.
3).
Fluke 435-II/437-II: check for a voltage readout V rms between 59.88 ... 60.12 V.
Table 4-3. Accuracy Check of Voltage Channels A/L1, B/L2, C/L3
Fluke 434-II: 59.4 ... 60.6 V
Fluke 435-II /437-II: 59.88 ... 60.12 V
Fluke 434-II: 119.4 ... 120.6 V
Fluke 435-II /437-II: 119.88 ... 120.12 V
Fluke 434-II: 239.4 ... 240.6 V
Fluke 435-II /437-II: 239.88 ... 240.12 V
Set the Calibrator to STBY.
(AMPL/FREQ key) to adjust the Calibrator to an Analyzer readout of 120.0 V /
120.00 V.
Fluke 434-II: check for a readout of 114.0 V or more.
Fluke 435-II/437-II: check for a readout of 114.00 V or more.
Check the channels according to the table below.
Table 4-4. Bandwidth Check of Voltage Channels A/L1, B/L2, C/L3
Fluke 434-II: 120.0 V
Fluke 435-II /437-II: 120.00 V
Fluke 434-II: 120.0 V
Fluke 435-II /437-II: 120.00 V
Fluke 434-II: 120.0 V
Fluke 435-II /437-II: 120.00 V
Readout at Voltage Channels
Readout at 60 Hz (adjust
Calibrator with EDIT FIELD)
Readout at 3 kHz
Fluke 434-II: ≥ 114.0 V
Fluke 435-II /437-II: ≥ 114.00 V
Fluke 434-II: ≥ 114.0 V
Fluke 435-II /437-II: ≥ 114.00 V
Fluke 434-II: ≥ 114.0 V
Fluke 435-II /437-II: ≥ 114.00 V